A comparison of two different mesh kit systems for anterior compartment prolapse repair.

Vivien Wong1, Ka Lai Shek, Ajay Rane, Joseph Lee, Anna Rosamilia, Hans Peter Dietz.   

Abstract

BACKGROUND AND AIM: Mesh reinforcement is considered an effective method for anterior compartment prolapse repair. Two common methods of mesh reinforcement involve either transobturator fixation (eg Perigee™) or lateral and apical anchoring (eg Anterior Elevate™). The aim of this study was to assess subjective and objective outcomes after Anterior Elevate and Perigee mesh kit surgery.
MATERIALS AND METHODS: This was a surgical audit of patients after anterior colporrhaphy (AC) with mesh reinforcement, undertaken at three tertiary urogynaecological centres. All patients were assessed for prolapse recurrence, which was defined as either (i) symptoms of prolapse (vaginal lump/dragging), (ii) ICS POPQ ≥ Stage 2, or (iii) bladder descent ≥10 mm below the symphysis pubis on transperineal ultrasound. Mesh co-ordinates and organ descent on Valsalva were determined relative to the inferior symphyseal margin.
RESULTS: Two hundred and twenty-nine patients with anterior compartment mesh (138 Perigee, 91 Elevate) were assessed at a median follow-up of 1.09 years (IQR 0.65-2.01). On assessment, 24% (n = 55) had symptoms of prolapse recurrence, 46% (n = 106) had a clinical recurrence, and 41% (n = 95) a recurrent cystocele sonographically. All objective results favoured the Perigee group. The superiority of the Perigee kit remained highly significant (P < 0.0001 for all clinical and ultrasound measures of prolapse recurrence) on multivariate analysis.
CONCLUSIONS: This retrospective analysis suggests that apical anchoring such as Anterior Elevate mesh system does not necessarily confer an advantage over the original transobturator mesh fixation technique for anterior compartment reconstruction.
© 2014 The Royal Australian and New Zealand College of Obstetricians and Gynaecologists.

Entities:  

Keywords:  3D ultrasound; anterior colporrhaphy; cystocele; mesh; pelvic organ prolapse; recurrence

Mesh:

Year:  2014        PMID: 24576207     DOI: 10.1111/ajo.12194

Source DB:  PubMed          Journal:  Aust N Z J Obstet Gynaecol        ISSN: 0004-8666            Impact factor:   2.100


  3 in total

Review 1.  Recurrent pelvic organ prolapse: International Urogynecological Association Research and Development Committee opinion.

Authors:  Sharif Ismail; Jonathan Duckett; Diaa Rizk; Olanrewaju Sorinola; Dorothy Kammerer-Doak; Oscar Contreras-Ortiz; Hazem Al-Mandeel; Kamil Svabik; Mitesh Parekh; Christian Phillips
Journal:  Int Urogynecol J       Date:  2016-07-05       Impact factor: 2.894

2.  Anterior-apical single-incision mesh surgery (uphold): 1-year outcomes on lower urinary tract symptoms, anatomy and ultrasonography.

Authors:  Tsia-Shu Lo; Leng Boi Pue; Yiap Loong Tan; Wu-Chiao Hsieh; Chuan Chi Kao; Ma Clarissa Uy-Patrimonio
Journal:  Int Urogynecol J       Date:  2018-07-14       Impact factor: 2.894

3.  A long-term comparative study of Uphold™ transvaginal mesh kit against anterior colporrhaphy.

Authors:  Moshe Gillor; Susanne Langer; Hans Peter Dietz
Journal:  Int Urogynecol J       Date:  2019-09-16       Impact factor: 2.894