Literature DB >> 2457559

A system for the automated data-acquisition of fast transient signals in excitable membranes.

J O Bustamante1.   

Abstract

This paper provides a description of a system for the acquisition of fast transient currents flowing across excitable membranes. The front end of the system consists of a CAMAC crate with plug-in modules. The modules provide control of CAMAC operations, analog to digital conversion, electronic memory storage and timing of events. The signals are transferred under direct memory access to an IBM PC microcomputer through a special-purpose interface. Voltage levels from a digital to analog board in the microcomputer are passed through multiplexers to produce the desired voltage pulse patterns to elicit the transmembrane currents. The dead time between consecutive excitatory voltage pulses is limited only by the computer data bus and the software characteristics. The dead time between data transfers can be reduced to the order of milliseconds, which is sufficient for most experiments with transmembrane ionic currents.

Mesh:

Substances:

Year:  1988        PMID: 2457559     DOI: 10.1016/0020-7101(88)90082-7

Source DB:  PubMed          Journal:  Int J Biomed Comput        ISSN: 0020-7101


  2 in total

1.  Restricted ion flow at the nuclear envelope of cardiac myocytes.

Authors:  J O Bustamante
Journal:  Biophys J       Date:  1993-06       Impact factor: 4.033

2.  Nuclear ion channels in cardiac myocytes.

Authors:  J O Bustamante
Journal:  Pflugers Arch       Date:  1992-08       Impact factor: 3.657

  2 in total