| Literature DB >> 24575279 |
A Elikaei1, S M Hosseini2, Z Sharifi1, H Latifi3, H Nikbakht3, H Mirshafiee2, A Asadollahi3.
Abstract
BACKGROUND: Despite improvement in safety of plasma transfusion some virus transmission still remains a problem. So as World Health Organization (WHO) recommends, many countries developed Pathogen Reduction Technologies (PRT) to inactivate pathogens, in plasma components. The Methylene Blue (MB) based methods is one of the most universal one. The purpose of this research was, produce a device that can inactivate viruses in MB environment.Entities:
Keywords: Blood-borne Pathogens; Methylene Blue; Plasma
Year: 2013 PMID: 24575279 PMCID: PMC3921873
Source DB: PubMed Journal: Iran J Ped Hematol Oncol ISSN: 2008-8892
Inactivation of HSV and VSV in human plasma by combination of 10 µM MB and visible light. Significant difference with initial load(P-value< 0.05).
|
|
|
|
| ||
|---|---|---|---|---|---|
|
|
|
|
| ||
|
| Initial Load | 6.70±0.10 | 6.20±0.03 | ||
|
| 0 J/cm2 | 6.50±0.12 | 6.00±0.05 | 0.20 | 0.20 |
|
| 27 J/cm2 | <3.10 | <3.20 | 4.60 | 3.00 |
|
| 40 J/cm2 | <1.15 | <1.5 | 5.55 | 4.70 |
|
| 54 J/cm2 | <0.5 | <0.7 | 6.20 | 5.50 |
|
| 70 J/cm2 | <0.5 | <0.5 | 6.20 | 5.30 |
Figure1Inactivation of HSV by 1 µM MB in combination with different light dose (depends on different illumination time), the most reduction obtained in 54 J/cm2 and increase in light dose decrease virus titration
Figure2Inactivation of VSV by1 µM MB in combination with different light dose (depends on different illumination time), the most reduction obtained in 70 J/cm2 and increase in light dose decrease virus titration
Inactivation of HSV, VSV in human plasma by combination of 1 µM MB and visible light. Significant Difference with initial load (P-value < 0.05).
|
|
|
|
| ||
|---|---|---|---|---|---|
| HSV | VSV | HSV | VSV | ||
|
| Initial Load | 6.70±0.10 | 6.20±0.03 | ||
|
| 0 J/cm2 | 6.50±0.12 | 6.00±0.05 | 0.20 | 0.20 |
|
| 27 J/cm2 | <4.90 | <4.70 | 1.80 | 1.50 |
|
| 40 J/cm2 | <2.50 | <3.00 | 4.20 | 3.20 |
|
| 54 J/cm2 | <1.50 | <1.90 | 5.20 | 4.30 |
|
| 70 J/cm2 | <1.50 | <1.30 | 5.20 | 4.90 |