Accuracy of tidal breathing measurement of FloRight compared to an ultrasonic flowmeter in infants.

Nicole C M Petrus1, Cindy Thamrin, Oliver Fuchs, Urs Frey.   

Abstract

INTRODUCTION: Monitoring breathing pattern is especially relevant in infants with lung disease. Recently, a vest-based inductive plethysmograph system (FloRight®) has been developed for tidal breathing measurement in infants. We investigated the accuracy of tidal breathing flow volume loop (TBFVL) measurements in healthy term-born infants and infants with lung disease by the vest-based system in comparison to an ultrasonic flowmeter (USFM) with a face mask. We also investigated whether the system discriminates between healthy infants and those with lung disease.
METHODS: Floright® measures changes in thoracoabdominal volume during tidal breathing through magnetic field changes generated by current-carrying conductor coils in an elastic vest. Simultaneous TBFVL measurements by the vest-based system and the USFM were performed at 44 weeks corrected postmenstrual age during quiet unsedated sleep. TBFVL parameters derived by both techniques and within both groups were compared.
RESULTS: We included 19 healthy infants and 18 infants with lung disease. Tidal volume per body weight derived by the vest-based system was significantly lower with a mean difference (95% CI) of -1.33 ml/kg (-1.73; -0.92), P < 0.001. Respiratory rate and ratio of time to peak tidal expiratory flow over total expiratory time (tPTEF/tE) did not differ between the two techniques. Both systems were able to discriminate between healthy infants and those with lung disease using tPTEF/tE.
CONCLUSION: FloRight® accurately measures time indices and may discriminate between healthy infants and those with lung disease, but demonstrates differences in tidal volume measurements. It may be better suited to monitor breathing pattern than for TBFVL measurements.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  infants; lung disease; lung function; validation

Mesh:

Year:  2014        PMID: 24574092     DOI: 10.1002/ppul.23006

Source DB:  PubMed          Journal:  Pediatr Pulmonol        ISSN: 1099-0496


  2 in total

1.  Electromagnetic inductance plethysmography is well suited to measure tidal breathing in infants.

Authors:  Mariann H L Bentsen; Morten Eriksen; Merete S Olsen; Trond Markestad; Thomas Halvorsen
Journal:  ERJ Open Res       Date:  2016-11-24

2.  Lung function at term in extremely preterm-born infants: a regional prospective cohort study.

Authors:  Mariann Haavik Bentsen; Trond Markestad; Knut Øymar; Thomas Halvorsen
Journal:  BMJ Open       Date:  2017-10-25       Impact factor: 2.692

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.