Literature DB >> 24566379

Understanding mechanisms and factors related to implant fixation; a model study of removal torque.

Patrik Stenlund1, Kohei Murase2, Christina Stålhandske3, Jukka Lausmaa4, Anders Palmquist5.   

Abstract

Osseointegration is a prerequisite for achieving a stable long-term fixation and load-bearing capacity of bone anchored implants. Removal torque measurements are often used experimentally to evaluate the fixation of osseointegrated screw-shaped implants. However, a detailed understanding of the way different factors influence the result of removal torque measurements is lacking. The present study aims to identify the main factors contributing to anchorage. Individual factors important for implant fixation were identified using a model system with an experimental design in which cylindrical or screw-shaped samples were embedded in thermosetting polymers, in order to eliminate biological variation. Within the limits of the present study, it is concluded that surface topography and the mechanical properties of the medium surrounding the implant affect the maximum removal torque. In addition to displaying effects individually, these factors demonstrate interplay between them. The rotational speed was found not to influence the removal torque measurements within the investigated range.
Copyright © 2014 Elsevier Ltd. All rights reserved.

Keywords:  Implant fixation; Removal torque; Surface roughness; Titanium implant

Mesh:

Substances:

Year:  2014        PMID: 24566379     DOI: 10.1016/j.jmbbm.2014.02.006

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  7 in total

1.  Temporal Changes in Reverse Torque of Locking-Head Screws Used in the Locking Plate in Segmental Tibial Defect in Goat Model.

Authors:  Remigiusz M Grzeskowiak; Rebecca E Rifkin; Elizabeth G Croy; Richard C Steiner; Reza Seddighi; Pierre-Yves Mulon; Henry S Adair; David E Anderson
Journal:  Front Surg       Date:  2021-04-27

2.  Osseointegration of Titanium Implants in Onlay of Cerament™, a New Ceramic Bone Substitute.

Authors:  Anna Truedsson; Jian-Sheng Wang; Pia Lindberg; Gunnar Warfvinge
Journal:  J Funct Biomater       Date:  2016-01-07

3.  Biomechanical evaluation of peak reverse torque (PRT) in a dynamic compression plate-screw construct used in a goat tibia segmental defect model.

Authors:  Remigiusz M Grzeskowiak; Carrie Wheeler; Elizabeth Taylor; James Lillich; James Roush; Alexandru S Biris; David E Anderson
Journal:  BMC Vet Res       Date:  2019-09-05       Impact factor: 2.741

4.  Effect of cyclic loading on the stability of screws placed in the locking plates used to bridge segmental bone defects.

Authors:  Remigiusz M Grzeskowiak; Laura R Freeman; David P Harper; David E Anderson; Pierre-Yves Mulon
Journal:  J Orthop Res       Date:  2020-09-09       Impact factor: 3.494

5.  Direct communication between osteocytes and acid-etched titanium implants with a sub-micron topography.

Authors:  Furqan A Shah; Patrik Stenlund; Anna Martinelli; Peter Thomsen; Anders Palmquist
Journal:  J Mater Sci Mater Med       Date:  2016-10-03       Impact factor: 3.896

6.  Three-dimensional modeling of removal torque and fracture progression around implants.

Authors:  Kohei Murase; Patrik Stenlund; Peter Thomsen; Jukka Lausmaa; Anders Palmquist
Journal:  J Mater Sci Mater Med       Date:  2018-06-30       Impact factor: 3.896

7.  Variation in bone response to the placement of percutaneous osseointegrated endoprostheses: A 24-month follow-up in sheep.

Authors:  Sujee Jeyapalina; James Peter Beck; Alex Drew; Roy D Bloebaum; Kent N Bachus
Journal:  PLoS One       Date:  2019-10-25       Impact factor: 3.240

  7 in total