| Literature DB >> 24561402 |
María-José De la Haba1, Mar Arias2, Pilar Ramírez3, María-Isabel López4, María-Teresa Sánchez5.
Abstract
This study assessed the potential of near infrared (NIR) spectroscopy as a non-destructive method for characterizing Protected Designation of Origin (PDO) "Vinagres de Montilla-Moriles" wine vinegars and for classifying them as a function of the manufacturing process used. Three spectrophotometers were evaluated for this purpose: two monochromator instruments (Foss NIRSystems 6500 SY-I and Foss NIRSystems 6500 SY-II; spectral range 400-2,500 nm in both cases) and a diode-array instrument (Corona 45 VIS/NIR; spectral range 380-1,700 nm). A total of 70 samples were used to predict major chemical quality parameters (total acidity, fixed acidity, volatile acidity, pH, dry extract, ash, acetoin, methanol, total polyphenols, color (tonality and intensity), and alcohol content), and to construct models for the classification of vinegars as a function of the manufacturing method used. The results obtained indicate that this non-invasive technology can be used successfully by the vinegar industry and by PDO regulators for the routine analysis of vinegars in order to authenticate them and to detect potential fraud. Slightly better results were achieved with the two monochromator instruments. The findings also highlight the potential of these NIR instruments for predicting the manufacturing process used, this being of particular value for the industrial authentication of traditional wine vinegars.Entities:
Year: 2014 PMID: 24561402 PMCID: PMC3958243 DOI: 10.3390/s140203528
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Basic technical characteristics of three spectrophotometers: two monochromators (M) and a diode-array instrument (DA).
|
| ||
|---|---|---|
| Detector type | Silicon, 400–1,100 nm. Lead Sulphide, 1,100–2,500 nm | Silicon, 400–950 nm. Indium-Gallium-Arsenide detector, 950–1,700 nm |
| Wavelength range (nm) | 400–2,500 | 380–1,700 |
| Spectral data rate | 1.8 scans/s | 20 scans/s |
| Dispersion | Pre | Post |
| Light source | Full spectrum | Full spectrum |
| Analysis mode | Reflectance | Reflectance |
Statistical analysis of calibration sample sets, i.e., data ranges, means and standard deviations (SD) and coefficients of variation (CV).
| Total acidity (g acetic acid/100 mL vinegar) | 70 | 4.65–15.75 | 9.44 | 2.14 | 22.70 |
| Fixed acidity (g acetic acid/100 mL vinegar) | 70 | 0.01–0.53 | 0.26 | 0.09 | 34.62 |
| Volatile acidity (g acetic acid/100 mL vinegar) | 70 | 4.39–14.81 | 9.19 | 2.10 | 22.87 |
| pH | 70 | 2.49–3.33 | 2.82 | 0.15 | 5.15 |
| Dry extract (g/L and percentage point of acetic acid) | 70 | 0.72–6.71 | 2.36 | 1.15 | 48.81 |
| Ash (g/L) | 70 | 1.40–8.25 | 3.32 | 1.42 | 42.68 |
| Acetoin (mg/L) | 70 | 144.75–2057.17 | 498.51 | 336.15 | 67.43 |
| Methanol (mg/L) | 70 | 12.18–122.74 | 49.85 | 21.28 | 42.69 |
| Total polyphenols (ppm gallic acid) | 70 | 73.39–801.05 | 344.29 | 141.08 | 40.98 |
| Intensity | 70 | 0.03–9.44 | 1.53 | 1.89 | 123.73 |
| Tonality | 70 | 0.10–4.53 | 2.56 | 1.30 | 50.88 |
| Alcohol content (% vol.) | 41 | 0.22–4.70 | 1.07 | 0.99 | 92.04 |
Figure 1.Typical log (1/R) spectra for Montilla-Moriles PDO vinegars.
Calibration statistics for quality parameters in wine vinegars.
| Total acidity (g acetic acid/100 mL vinegar) | FNS I | 2,5,5,1 | SNV + DT | 400–2500 | 9.49 | 2.07 | 0.25 | 0.99 | 8.35 | 2.62 |
| FNS II | 1,5,5,1 | None | 400–2500 | 9.50 | 2.08 | 0.26 | 0.98 | 7.95 | 2.76 | |
| Corona | 1,5,5,1 | None | 500–1690 | 9.39 | 2.17 | 0.29 | 0.98 | 7.44 | 3.12 | |
| Fixed acidity (g acetic acid/100 mL vinegar) | FNS I | 1,10,5,1 | SNV + DT | 1100–2500 | 0.24 | 0.08 | 0.04 | 0.79 | 2.19 | 15.62 |
| FNS II | 1,10,5,1 | SNV + DT | 1100–2500 | 0.25 | 0.09 | 0.04 | 0.78 | 2.15 | 16.70 | |
| Corona | 2,5,5,1 | SNV + DT | 500–1690 | 0.25 | 0.08 | 0.05 | 0.58 | 1.53 | 19.77 | |
| Volatile acidity (g acetic acid/100 mL vinegar) | FNS I | 2,5,5,1 | SNV + DT | 400–2500 | 9.24 | 2.03 | 0.25 | 0.98 | 7.99 | 2.76 |
| FNS II | 1,5,5,1 | None | 400–2500 | 9.11 | 2.10 | 0.29 | 0.98 | 7.16 | 3.22 | |
| Corona | 1,5,5,1 | None | 500–1690 | 9.10 | 2.13 | 0.31 | 0.98 | 6.92 | 3.38 | |
| pH | FNS I | 1,5,5,1 | None | 1100–2500 | 2.82 | 0.14 | 0.05 | 0.85 | 2.60 | 1.85 |
| FNS II | 1,5,5,1 | None | 400–2500 | 2.81 | 0.12 | 0.05 | 0.81 | 2.25 | 1.95 | |
| Corona | 1,5,5,1 | None | 500–1690 | 2.80 | 0.13 | 0.07 | 0.72 | 1.89 | 2.43 | |
| Dry extract (g/L and percentage point of acetic acid) | FNS I | 2,10,5,1 | SNV + DT | 400–2500 | 2.33 | 1.13 | 0.14 | 0.99 | 8.16 | 5.94 |
| FNS II | 1,5,5,1 | SNV + DT | 1100–2500 | 2.32 | 1.12 | 0.12 | 0.99 | 9.41 | 5.14 | |
| Corona | 1,5,5,1 | None | 1100–1690 | 2.10 | 0.75 | 0.18 | 0.94 | 4.14 | 8.65 | |
| Ash (g/L) | FNS I | 1,5,5,1 | SNV + DT | 1100–2500 | 3.26 | 1.33 | 0.41 | 0.91 | 3.28 | 12.49 |
| FNS II | 1,10,5,1 | SNV + DT | 1100–2500 | 3.28 | 1.37 | 0.30 | 0.95 | 4.59 | 9.10 | |
| Corona | 2,10,5,1 | SNV + DT | 500–1690 | 3.29 | 1.35 | 0.55 | 0.83 | 1.62 | 16.71 | |
| Acetoin (mg/L) | FNS I | 2,5,5,1 | SNV + DT | 400–2500 | 439.26 | 237.45 | 127.33 | 0.71 | 1.86 | 28.99 |
| FNS II | 1,10,5,1 | SNV + DT | 400–2500 | 436.34 | 239.58 | 158.82 | 0.56 | 1.51 | 36.40 | |
| Corona | 2,10,5,1 | None | 500–1690 | 459.69 | 244.56 | 155.92 | 0.59 | 1.56 | 33.92 | |
| Methanol (mg/L) | FNS I | 2,10,5,1 | None | 1100–2500 | 46.94 | 17.04 | 9.85 | 0.67 | 1.73 | 20.98 |
| FNS II | 1,5,5,1 | SNV + DT | 400–2500 | 46.64 | 17.25 | 7.81 | 0.80 | 2.21 | 16.74 | |
| Corona | 2,5,5,1 | None | 500–1690 | 44.35 | 14.67 | 7.93 | 0.71 | 1.85 | 17.87 | |
| Total polyphenols (ppm gallic acid) | FNS I | 1,5,5,1 | SNV + DT | 400–2500 | 337.08 | 133.35 | 56.69 | 0.82 | 2.35 | 16.82 |
| FNS II | 1,10,5,1 | None | 400–2500 | 337.08 | 133.35 | 58.42 | 0.81 | 2.28 | 17.33 | |
| Corona | 2,5,5,1 | SNV + DT | 500–1690 | 337.99 | 136.33 | 63.58 | 0.78 | 2.14 | 18.81 | |
| Intensity | FNS I | 1,10,5,1 | SNV + DT | 400–2500 | 1.31 | 1.55 | 0.31 | 0.96 | 4.96 | 23.93 |
| FNS II | 2,10,5,1 | SNV + DT | 400–2500 | 1.48 | 1.66 | 0.23 | 0.98 | 6.86 | 16.41 | |
| Corona | 1,5,5,1 | SNV +DT | 500–1690 | 1.31 | 1.56 | 0.27 | 0.97 | 5.88 | 20.32 | |
| Tonality | FNS I | 2,10,5,1 | SNV + DT | 400–2500 | 2.54 | 1.31 | 0.49 | 0.86 | 2.68 | 19.30 |
| FNS II | 2,5,5,1 | None | 400–2500 | 2.60 | 1.28 | 0.41 | 0.90 | 3.13 | 15.79 | |
| Corona | 1,5,5,1 | SNV + DT | 500–1690 | 2.67 | 1.25 | 0.30 | 0.94 | 4.19 | 11.14 | |
| Alcohol content (% vol.) | FNS I | 2,5,5,1 | SNV + DT | 400–2500 | 0.98 | 0.81 | 0.06 | 0.99 | 14.42 | 5.71 |
| FNS II | 2,10,5,1 | None | 400–2500 | 0.96 | 0.81 | 0.04 | 0.99 | 22.31 | 3.76 | |
| Corona | 1,5,5,1 | SNV + DT | 1100–1690 | 1.09 | 0.99 | 0.20 | 0.96 | 4.98 | 18.21 | |
Note:
1 Mean of the calibration set;
2 Standard deviation;
3 Standard error of cross-validation;
4 Coefficient of determination of cross-validation;
5 Ratio SD/SECV;
6 Coefficient of variation;
The best of the best equations for each parameter and instrument tested.
Percentage of Montilla-Moriles PDO vinegars classified by manufacturing method. PLS-DA.
|
| ||||||
|---|---|---|---|---|---|---|
| A: 94.3% | A: 92.8% | A: 94.3% | A: 91.7% | A: 93.7% | A: 93.7% | |
| B: 8 | B: 8 | B: 6 | B: 5 | B: 6 | B: 4 | |
| C: 1,5,5,1 | C: 1,5,5,1 | C: 1,5,5,1 | C: 1,5,5,1 | C: 2,10,5,1 | C: 2,5,5,1 | |
|
| ||||||
|
| ||||||
| Traditional | 87.5% | 91.7% | 91.6% | 87.5% | 95.8% | 91.7% |
| Submerged | 97.8% | 93.5% | 95.6% | 95.8% | 91.7% | 95.8% |
Note:
A Percentage of correctly classified training samples after cross validation;
B Number of synthetic variables;
C Math treatment;
* The best of the best models for the instruments studied.
Figure 2.D1 log (1/R) spectra for Montilla-Moriles PDO vinegars made by traditional and submerged methods. Corona 45 VIS/NIR spectrophotometer. Spectral range 500–1,690 nm.