Literature DB >> 24560093

A new cochlear implant electrode with a "cork"-type stopper for inner ear malformations.

Levent Sennaroğlu1, Gamze Atay1, Münir Demir Bajin2.   

Abstract

OBJECTIVE: Gusher in inner ear malformations is common in patients with incomplete partition type I and type III. It is also common in less severe form as oozing in incomplete partition type II and large vestibular aqueduct. It is important to prevent cerebrospinal fluid (CSF) escape around the electrode to prevent meningitis.
METHODS: The custom-made device was produced by Med-El Company. It has a "cork"-like stopper instead of the usual silicon ring to prevent gusher. There are two types of electrodes of different lengths. The standard one is 25mm (contact space 1.7mm) and the short one is 20mm (contact space 1.3mm). It was used in 50 patients with different inner ear malformations.
RESULTS: Thirteen patients had gusher, and 11 patients oozing during cochleostomy. One patient with initial prototype of the cork electrode had to be revised because of persistent oozing around the electrode. Another patient had slow extrusion of the electrode most probably due to CSF pulsation and had to be revised. Both patients had no more CSF fistula.
CONCLUSION: CSF fistula in inner ear malformations is a serious situation which may lead to recurrent meningitis. The new electrode with "cork" stopper looks promising in preventing the postoperative CSF leak around the electrode.
Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  Cochlear implantation; Gusher; Inner ear anomaly

Mesh:

Year:  2014        PMID: 24560093     DOI: 10.1016/j.anl.2013.12.011

Source DB:  PubMed          Journal:  Auris Nasus Larynx        ISSN: 0385-8146            Impact factor:   1.863


  6 in total

Review 1.  Cochlear implantation in adults and pediatrics with enlarged vestibular aqueduct: a systematic review on the surgical findings and patients' performance.

Authors:  Asma Alahmadi; Yassin Abdelsamad; Marzouqi Salamah; Saad Alenzi; Khalid M Badr; Saeed Alghamdi; Abdulrahman Alsanosi
Journal:  Eur Arch Otorhinolaryngol       Date:  2022-06-30       Impact factor: 2.503

2.  A New Pathogenic Variant in POU3F4 Causing Deafness Due to an Incomplete Partition of the Cochlea Paved the Way for Innovative Surgery.

Authors:  Ahmet M Tekin; Marco Matulic; Wim Wuyts; Masoud Zoka Assadi; Griet Mertens; Vincent van Rompaey; Yongxin Li; Paul van de Heyning; Vedat Topsakal
Journal:  Genes (Basel)       Date:  2021-04-21       Impact factor: 4.096

Review 3.  Cochlear implantation in common cavity deformity: a systematic review.

Authors:  Ayshah Al-Mahboob; Salman F Alhabib; Yassin Abdelsamad; Farid Alzhrani
Journal:  Eur Arch Otorhinolaryngol       Date:  2021-05-29       Impact factor: 2.503

4.  Intracochlear fluid pressure changes related to the insertional speed of a CI electrode.

Authors:  I Todt; P Mittmann; A Ernst
Journal:  Biomed Res Int       Date:  2014-07-16       Impact factor: 3.411

5.  Classification and Current Management of Inner Ear Malformations.

Authors:  Levent Sennaroğlu; Münir Demir Bajin
Journal:  Balkan Med J       Date:  2017-08-25       Impact factor: 2.021

6.  Analysis of Long-Term Cochlear Implantation Outcomes and Correlation With Imaging Characteristics in Patients With Common Cavity Deformity.

Authors:  Lifang Zhang; Biao Chen; Ying Kong; Natalia Liau; Xingmei Wei; Ying Shi; Jingyuan Chen; Mengge Yang; Anandhan Dhanasingh; Yongxin Li
Journal:  Front Neurosci       Date:  2022-03-23       Impact factor: 4.677

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.