Literature DB >> 24556873

Real-time proton beam range monitoring by means of prompt-gamma detection with a collimated camera.

F Roellinghoff1, A Benilov, D Dauvergne, G Dedes, N Freud, G Janssens, J Krimmer, J M Létang, M Pinto, D Prieels, C Ray, J Smeets, F Stichelbaut, E Testa.   

Abstract

Prompt-gamma profile was measured at WPE-Essen using 160 MeV protons impinging a movable PMMA target. A single collimated detector was used with time-of-flight (TOF) to reduce the background due to neutrons. The target entrance rise and the Bragg peak falloff retrieval precision was determined as a function of incident proton number by a fitting procedure using independent data sets. Assuming improved sensitivity of this camera design by using a greater number of detectors, retrieval precisions of 1 to 2 mm (rms) are expected for a clinical pencil beam. TOF improves the contrast-to-noise ratio and the performance of the method significantly.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24556873     DOI: 10.1088/0031-9155/59/5/1327

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  6 in total

Review 1.  In vivo range verification in particle therapy.

Authors:  Katia Parodi; Jerimy C Polf
Journal:  Med Phys       Date:  2018-11       Impact factor: 4.071

2.  3D prompt gamma imaging for proton beam range verification.

Authors:  E Draeger; D Mackin; S Peterson; H Chen; S Avery; S Beddar; J C Polf
Journal:  Phys Med Biol       Date:  2018-01-30       Impact factor: 3.609

3.  First Images of a Three-Layer Compton Telescope Prototype for Treatment Monitoring in Hadron Therapy.

Authors:  Gabriela Llosá; Marco Trovato; John Barrio; Ane Etxebeste; Enrique Muñoz; Carlos Lacasta; Josep F Oliver; Magdalena Rafecas; Carles Solaz; Paola Solevi
Journal:  Front Oncol       Date:  2016-02-02       Impact factor: 6.244

4.  Assessment of Geant4 Prompt-Gamma Emission Yields in the Context of Proton Therapy Monitoring.

Authors:  Marco Pinto; Denis Dauvergne; Nicolas Freud; Jochen Krimmer; Jean M Létang; Etienne Testa
Journal:  Front Oncol       Date:  2016-01-28       Impact factor: 6.244

Review 5.  Range Verification Methods in Particle Therapy: Underlying Physics and Monte Carlo Modeling.

Authors:  Aafke Christine Kraan
Journal:  Front Oncol       Date:  2015-07-07       Impact factor: 6.244

6.  Experimental Comparison of Knife-Edge and Multi-Parallel Slit Collimators for Prompt Gamma Imaging of Proton Pencil Beams.

Authors:  Julien Smeets; Frauke Roellinghoff; Guillaume Janssens; Irene Perali; Andrea Celani; Carlo Fiorini; Nicolas Freud; Etienne Testa; Damien Prieels
Journal:  Front Oncol       Date:  2016-06-27       Impact factor: 6.244

  6 in total