Literature DB >> 24519902

Stent fracture: Insights on mechanisms, treatments, and outcomes from the food and drug administration manufacturer and user facility device experience database.

Mamas A Mamas1, Nicolas Foin, Chad Abunassar, Muhammad A Khan, Carlo Di Mario, Douglas G Fraser.   

Abstract

OBJECTIVES: Stent fracture (SF) is an uncommon complication following percutaneous coronary intervention. Previous studies of SF have either been small single-center studies or have reported data mainly from first generation platforms. The FDA MAUDE database was used to identify cases of SF to gain mechanistic insight into procedural and anatomical factors predisposing to this complication, to define treatment and associated clinical outcomes in contemporary and first generation stent platforms. METHODS AND
RESULTS: Using defined search terms, 376 unique reports of SF were identified in the FDA MAUDE database of which 336 (89.4%) occurred in first generation stent platforms with 96.1% occurring in the Cypher platform. The remaining 40 reports (10.6%) occurred with a similar distribution of cases across different contemporary platforms. Common vessel characteristics associated with SF included calcification, tortuosity, stent overlap, and significant angulation with the most common modality of presentation being in-stent restenosis or stent thrombosis.
CONCLUSIONS: SF occurs most commonly in first generation Cypher stents with a similar distribution of cases across different contemporary platforms. Optimal treatment strategy remains unclear with the decision to treat and the modality of treatment utilized depending on the type and severity of SF.
Copyright © 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  food and drug administration; maude database; stent design/structure/coatings; stent fracture/failure; stent restenosis; stent thrombosis

Mesh:

Year:  2014        PMID: 24519902     DOI: 10.1002/ccd.25423

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  2 in total

Review 1.  Influences of Stent Design on In-Stent Restenosis and Major Cardiac Outcomes: A Scoping Review and Meta-Analysis.

Authors:  Omer Burak Istanbullu; Gulsen Akdogan
Journal:  Cardiovasc Eng Technol       Date:  2021-08-18       Impact factor: 2.495

2.  Thoracic Endovascular Aortic Repair Adverse Events Reported In The Food And Drug Administration Manufacturer And User Facility Device Experience Database.

Authors:  Neel A Mansukhani; Meraaj S Haleem; Mark K Eskandari
Journal:  Med Devices (Auckl)       Date:  2019-11-14