Reproducibility of disc and macula optical coherence tomography using the Canon OCT-HS100 as compared with the Zeiss Cirrus HD-OCT.

Rune Brautaset1, Ulrika Birkeldh, Rebecka Rosén, Marika Wahlberg Ramsay, Maria Nilsson.   

Abstract

PURPOSE: In a clinical setting, the usefulness of optical coherence tomography (OCT) is strongly dependent on reproducibility of the measurement. The aim of the present study was to evaluate macula and optic disc measurement reproducibility with the new spectral-domain OCT (SD-OCT) from Canon (Canon OCT-HS100) and to compare reproducibility and obtained measurements with the Zeiss Cirrus HD-OCT.
METHODS: Macula and optic disc parameters from the right eyes of 31 subjects were obtained twice with both instruments. Interoperator reproducibility was evaluated by use of the coefficient of repeatability (CR), and the obtained measurements were compared between the instruments.
RESULTS: No difference in interoperator reproducibility could be found when comparing the 2 instruments and reproducibility ranged from 3.94% to 12.77% for optic disc parameters and from 1.19% to 3.54% for macula parameters. The lowest reproducibility was found for cup volume and vertical cup/disc ratio with both instruments. For all macula and retinal nerve fiber layer (RNFL) thickness measurements, there was a statistical difference when comparing the 2 instruments, except for RFNL measurements of the superior quadrant, with the Canon OCT-HS100 always evaluating the thickness to be thicker; however, the 2 instruments correlated well.
CONCLUSIONS: The Canon OCT-HS100 is a reproducible instrument that matches the Zeiss Cirrus HD-OCT well. It remains to be evaluated how sensitive the Canon OCT-HS100 is to detect small subtle changes in optic disc parameters and macular nerve fiber layer thickness. Furthermore, due to the differences in thickness estimation, it is important to emphasize that SD-OCTs are not interchangeable.

Entities:  

Mesh:

Year:  2014        PMID: 24519507     DOI: 10.5301/ejo.5000437

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  4 in total

1.  Changes in Schlemm's canal, trabecular meshwork, and relevant parameters in the early stage after SMILE of myopia patients.

Authors:  Dan-Dan Yang; Zhi-Qi Chen; Wei Chen; He Yin; Jing-Kai Peng; Jun-Ming Wang
Journal:  Int J Ophthalmol       Date:  2022-02-18       Impact factor: 1.779

2.  Topographic distribution features of the choroidal and retinal nerve fiber layer thickness in Chinese school-aged children.

Authors:  Wei-Qin Liu; Dan-Dan Wang; Xiao-Xia Yang; Yan-Yan Pan; Xue Song; Yu-Shan Hou; Chen-Xiao Wang
Journal:  Int J Ophthalmol       Date:  2020-09-18       Impact factor: 1.779

3.  Repeatability Using Automatic Tracing with Canon OCT- HS100 and Zeiss Cirrus HD-OCT 5000.

Authors:  Rune Brautaset; Ulrika Birkeldh; Petra Frehr Alstig; Petra Wikén; Maria Nilsson
Journal:  PLoS One       Date:  2016-02-11       Impact factor: 3.240

4.  The Temporal Retinal Nerve Fiber Layer Thickness Is the Most Important Optical Coherence Tomography Estimate in Multiple Sclerosis.

Authors:  Ulrika Birkeldh; Ali Manouchehrinia; Max Albert Hietala; Jan Hillert; Tomas Olsson; Fredrik Piehl; Ingrid Skelton Kockum; Lou Brundin; Ori Zahavi; Marika Wahlberg-Ramsay; Rune Brautaset; Maria Nilsson
Journal:  Front Neurol       Date:  2017-12-13       Impact factor: 4.003

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.