In vitro comparison of five nebulizers during noninvasive ventilation: analysis of inhaled and lost doses.

Jean-Bernard Michotte1, Emilie Jossen, Jean Roeseler, Giuseppe Liistro, Grégory Reychler.   

Abstract

BACKGROUND: Few studies on performance comparison of nebulizer systems coupled with a single-limb circuit bilevel ventilator are available. Most of these data compared the aerosol drug delivery for only two different systems. Using an adult lung bench model of noninvasive ventilation, we compared inhaled and lost doses of three nebulizer systems coupled with a single-limb circuit bilevel ventilator, as well as the influence of the nebulizer position.
METHOD: Three vibrating mesh nebulizers (Aeroneb(®) Pro, Aeroneb(®) Solo, and NIVO(®)), one jet nebulizer (Sidestream(®)), and one ultrasonic nebulizer (Servo Ultra Nebulizer 145(®)) coupled with a bilevel ventilator were tested. They were charged with amikacin solution (500 mg/4 mL) and operated at two different positions: before and after the exhalation port (starting from the lung). The inhaled dose, the expiratory wasted dose, and the estimated lost dose were assessed by the residual gravimetric method.
RESULTS: The doses varied widely among the nebulizer types and position. When the nebulizer was positioned before the exhalation port, the vibrating mesh nebulizer delivered the highest inhaled dose (p<0.001), the jet nebulizer the highest expiratory wasted dose (p<0.001), and the ultrasonic device the highest total lost dose (p<0.001). When the nebulizer was positioned after the exhalation port, the vibrating mesh nebulizers delivered the highest inhaled (p<0.001) and expiratory wasted doses (p<0.001), and the ultrasonic device the highest total lost dose (p<0.001). The most efficient nebulizers were NIVO and Aeroneb Solo when placed before the exhalation port.
CONCLUSIONS: In a single-limb circuit bilevel ventilator, vibrating mesh nebulizers positioned between the exhalation port and lung model are more efficient for drug delivery compared with jet or ultrasonic nebulizers. In this position, the improved efficiency of vibrating mesh nebulizers was due to an increase in the inhaled dose and a reduction in the exhaled wasted dose compared with placement between the ventilator and the expiratory port. Because of the high total lost dose, the ultrasonic device should not be recommended. Nebulizer placement before the exhalation port increased the inhaled dose and decreased the expiratory wasted dose, except for the jet nebulizer.

Entities:  

Keywords:  aerosol delivery; bilevel ventilator; jet nebulizer; nebulizer; noninvasive ventilation; single-limb circuit ventilator; ultrasonic nebulizer; vibrating mesh nebulizer

Mesh:

Substances:

Year:  2014        PMID: 24517084     DOI: 10.1089/jamp.2013.1070

Source DB:  PubMed          Journal:  J Aerosol Med Pulm Drug Deliv        ISSN: 1941-2711            Impact factor:   2.849


  8 in total

Review 1.  Aerosol delivery via noninvasive ventilation: role of models and bioanalysis.

Authors:  Haitham Saeed; Hadeer S Harb; Yasmin M Madney; Mohamed E A Abdelrahim
Journal:  Ann Transl Med       Date:  2021-04

2.  The utilization of aerosol therapy in mechanical ventilation patients: a prospective multicenter observational cohort study and a review of the current evidence.

Authors:  Shan Lyu; Jie Li; Limin Yang; Xiaoliang Du; Xiaoyi Liu; Libo Chuan; Guoqiang Jing; Zhenyan Wang; Weiwei Shu; Chunjuan Ye; Qionglan Dong; Jun Duan; James B Fink; Zhancheng Gao; Zongan Liang
Journal:  Ann Transl Med       Date:  2020-09

3.  Effect of pressures and type of ventilation on aerosol delivery to chronic obstructive pulmonary disease patients.

Authors:  Marina E Boules; Nabila Ibrahim Laz; Ahmed A Elberry; Raghda R S Hussein; Mohamed E A Abdelrahim
Journal:  Beni Suef Univ J Basic Appl Sci       Date:  2022-04-15

Review 4.  Fundamentals of aerosol therapy in critical care.

Authors:  Jayesh Dhanani; John F Fraser; Hak-Kim Chan; Jordi Rello; Jeremy Cohen; Jason A Roberts
Journal:  Crit Care       Date:  2016-10-07       Impact factor: 9.097

Review 5.  Inhalation Techniques Used in Patients with Respiratory Failure Treated with Noninvasive Mechanical Ventilation.

Authors:  Patrycja Rzepka-Wrona; Szymon Skoczynski; Dawid Wrona; Adam Barczyk
Journal:  Can Respir J       Date:  2018-06-03       Impact factor: 2.409

6.  Aerosol Delivery to a Critically Ill Patient: A Big Issue Easily Solved by Developing Guidelines.

Authors:  Mohamed E A Abdelrahim
Journal:  Pulm Ther       Date:  2018-07-31

7.  Effect of IFN-α and other commonly used nebulization drugs in different nebulization methods on the resistance of breathing circuit filters under invasive mechanical ventilation.

Authors:  Zhenjie Jiang; Hanwen Liang; Guixia Peng; Shiya Wang; Baozhu Zhang; Qingwen Sun; Yuanda Xu; Huiqing Zeng; Jingye Huang
Journal:  Ann Transl Med       Date:  2022-02

8.  A case of severe acute exacerbation of Yokkaichi asthma treated with a vibrating mesh nebulizer.

Authors:  Takeshi Yano; Tetsu Yonaha; Koutaro Hidaka; Masumi Nagahama; Tomohiro Koshida; Hiroshi Matsuoka; Masahiko Taniguchi; Isao Tsuneyoshi
Journal:  Respir Med Case Rep       Date:  2016-08-03
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.