Literature DB >> 24512403

Inappropriate implantable cardioverter-defibrillator shocks attributed to alternating-current leak in a swimming pool.

John N Makaryus1, Julia Angert-Gilman, Mena Yacoub, Apoor Patel, Bruce Goldner.   

Abstract

Implantable cardioverter-defibrillators (ICDs) are the standard of care for preventing sudden cardiac death in patients who are predisposed to malignant ventricular arrhythmias. Causes of inappropriate ICD shock include equipment malfunction, improper arrhythmia evaluation, misinterpretation of myopotentials, and electromagnetic interference. As the number of implanted ICDs has increased, other contributors to inappropriate therapy have become known, such as minimal electrical current leaks that mimic ventricular fibrillation. We present the case of a 63-year-old man with a biventricular ICD who received 2 inappropriate shocks, probably attributable to alternating-current leaks in a swimming pool. In addition, we discuss ICD sensitivity and offer recommendations to avoid similar occurrences.

Entities:  

Mesh:

Year:  2014        PMID: 24512403      PMCID: PMC3967468          DOI: 10.14503/THIJ-12-3005

Source DB:  PubMed          Journal:  Tex Heart Inst J        ISSN: 0730-2347


  10 in total

1.  Inadvertent detection of 60-Hz alternating current by an implantable cardioverter defibrillator.

Authors:  Sung W Lee; Jeffrey P Moak; Brian Lewis
Journal:  Pacing Clin Electrophysiol       Date:  2002-04       Impact factor: 1.976

Review 2.  Recording techniques for clinical electrophysiology.

Authors:  William G Stevenson; Kyoko Soejima
Journal:  J Cardiovasc Electrophysiol       Date:  2005-09

3.  The response of implanted dual chamber pacemakers to 50 Hz extraneous electrical interference.

Authors:  P S Astridge; G C Kaye; S Whitworth; P Kelly; A J Camm; E J Perrins
Journal:  Pacing Clin Electrophysiol       Date:  1993-10       Impact factor: 1.976

4.  Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction.

Authors:  Arthur J Moss; Wojciech Zareba; W Jackson Hall; Helmut Klein; David J Wilber; David S Cannom; James P Daubert; Steven L Higgins; Mary W Brown; Mark L Andrews
Journal:  N Engl J Med       Date:  2002-03-19       Impact factor: 91.245

5.  Inappropriate implantable cardioverter-defibrillator shock due to external alternating current leak: report of two cases.

Authors:  Ngai Yin Chan; Lillian Wai-Ling Ho
Journal:  Europace       Date:  2005-03       Impact factor: 5.214

6.  True bipolar and integrated bipolar sensing and detection by implantable defibrillators.

Authors:  Trevor A Verga; Jeffrey M Gillberg; Richard M Greenberg; Florin T Deger
Journal:  Pacing Clin Electrophysiol       Date:  2011-11       Impact factor: 1.976

7.  Noise, artifact, and oversensing related inappropriate ICD shock evaluation: ALTITUDE noise study.

Authors:  Brian D Powell; Samuel J Asirvatham; David L Perschbacher; Paul W Jones; Yong-Mei Cha; David A Cesario; Michael Cao; F Roosevelt Gilliam; Leslie A Saxon
Journal:  Pacing Clin Electrophysiol       Date:  2012-04-22       Impact factor: 1.976

8.  Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact.

Authors:  James P Daubert; Wojciech Zareba; David S Cannom; Scott McNitt; Spencer Z Rosero; Paul Wang; Claudio Schuger; Jonathan S Steinberg; Steven L Higgins; David J Wilber; Helmut Klein; Mark L Andrews; W Jackson Hall; Arthur J Moss
Journal:  J Am Coll Cardiol       Date:  2008-04-08       Impact factor: 24.094

9.  Prophylactic defibrillator implantation in patients with nonischemic dilated cardiomyopathy.

Authors:  Alan Kadish; Alan Dyer; James P Daubert; Rebecca Quigg; N A Mark Estes; Kelley P Anderson; Hugh Calkins; David Hoch; Jeffrey Goldberger; Alaa Shalaby; William E Sanders; Andi Schaechter; Joseph H Levine
Journal:  N Engl J Med       Date:  2004-05-20       Impact factor: 91.245

10.  Inappropriate implantable cardioverter defibrillator discharge from sensing of external alternating current leak.

Authors:  Ashok Garg; Manish Wadhwa; Keith Brown; Cathy Luckett; Tim Vaughn; Ulrika Birgersdotter-Green; Gregory Feld
Journal:  J Interv Card Electrophysiol       Date:  2002-10       Impact factor: 1.900

  10 in total
  2 in total

1.  Electromagnetic Interference in a Private Swimming Pool: Case report.

Authors:  Sandia Iskandar; Madhav Lavu; Moustapha Atoui; Dhanunjaya Lakkireddy
Journal:  Indian Pacing Electrophysiol J       Date:  2016-02-12

2.  Swimming pool saline chlorination units and implantable cardiac devices: A source for potentially fatal electromagnetic interference.

Authors:  John Wight; Michael S Lloyd
Journal:  HeartRhythm Case Rep       Date:  2019-01-30