Literature DB >> 2451231

The porous titanium steroid eluting electrode: a double blind study assessing the stimulation threshold effects of steroid.

H Mond1, K Stokes, J Helland, L Grigg, P Kertes, B Pate, D Hunt.   

Abstract

A transvenous pacing lead with a porous electrode which slowly elutes the steroid, dexamethasone sodium phosphate, has been developed. Previous investigations show low and constant stimulation thresholds persisting over at least the first two years post-implantation. As it is not known whether this low threshold results from the steroid or electrode configuration, a double blind study was designed to compare the same electrode configuration with and without steroid over a 2-year follow-up period. There were ten patients in each group with similar age, sex, indications for pacing and implantation data. Regular measurements of postoperative pulse duration thresholds were performed using a customized VVIM pulse generator programmed to 1.5 V output. For the first two days post-implantation, there were no statistical differences in the pulse duration thresholds between the two pacing leads. From 2 weeks to 2 years the pulse duration thresholds for the steroid leads remained almost constant, whereas the leads without steroid showed a typical rise. The difference in pulse duration thresholds between the two groups of leads from two weeks onwards confirmed that it was the steroid rather than the electrode configuration which prevented the rise in chronic stimulation threshold.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 2451231     DOI: 10.1111/j.1540-8159.1988.tb04543.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  7 in total

1.  Pacing threshold increase in nonthoracotomy implantable defibrillator leads: implications for battery longevity and margin of safety.

Authors:  A E Epstein; V J Plumb; K A Kirk; G N Kay
Journal:  J Interv Card Electrophysiol       Date:  1997-09       Impact factor: 1.900

2.  The biocompatibility manifesto: biocompatibility for the twenty-first century.

Authors:  Buddy D Ratner
Journal:  J Cardiovasc Transl Res       Date:  2011-06-28       Impact factor: 4.132

3.  A Materials Roadmap to Functional Neural Interface Design.

Authors:  Steven M Wellman; James R Eles; Kip A Ludwig; John P Seymour; Nicholas J Michelson; William E McFadden; Alberto L Vazquez; Takashi D Y Kozai
Journal:  Adv Funct Mater       Date:  2017-07-19       Impact factor: 18.808

4.  Sick sinus syndrome and diffuse impairment of the conduction system in a child: successful pacing with a steroid eluting endocardial pacing lead.

Authors:  A Capucci; G Boriani; R Galli; F M Picchio; A Pierangeli; B Magnani
Journal:  Pediatr Cardiol       Date:  1992-01       Impact factor: 1.655

5.  Local delivery of dexamethasone for prevention of neointimal proliferation in a rat model of balloon angioplasty.

Authors:  A E Villa; L A Guzman; W Chen; G Golomb; R J Levy; E J Topol
Journal:  J Clin Invest       Date:  1994-03       Impact factor: 14.808

6.  Inhibition of Na+/H+exchanger modulates microglial activation and scar formation following microelectrode implantation.

Authors:  Mitchell Dubaniewicz; James R Eles; Stephanie Lam; Shanshan Song; Franca Cambi; Dandan Sun; Steven M Wellman; Takashi D Y Kozai
Journal:  J Neural Eng       Date:  2021-03-19       Impact factor: 5.379

7.  Wireless, battery-free, fully implantable multimodal and multisite pacemakers for applications in small animal models.

Authors:  Philipp Gutruf; Rose T Yin; K Benjamin Lee; Jokubas Ausra; Jaclyn A Brennan; Yun Qiao; Zhaoqian Xie; Roberto Peralta; Olivia Talarico; Alejandro Murillo; Sheena W Chen; John P Leshock; Chad R Haney; Emily A Waters; Changxing Zhang; Haiwen Luan; Yonggang Huang; Gregory Trachiotis; Igor R Efimov; John A Rogers
Journal:  Nat Commun       Date:  2019-12-17       Impact factor: 17.694

  7 in total