A methodological evaluation of volumetric measurement techniques including three-dimensional imaging in breast surgery.

H Hoeffelin1, D Jacquemin1, V Defaweux2, J L Nizet1.   

Abstract

Breast surgery currently remains very subjective and each intervention depends on the ability and experience of the operator. To date, no objective measurement of this anatomical region can codify surgery. In this light, we wanted to compare and validate a new technique for 3D scanning (LifeViz 3D) and its clinical application. We tested the use of the 3D LifeViz system (Quantificare) to perform volumetric calculations in various settings (in situ in cadaveric dissection, of control prostheses, and in clinical patients) and we compared this system to other techniques (CT scanning and Archimedes' principle) under the same conditions. We were able to identify the benefits (feasibility, safety, portability, and low patient stress) and limitations (underestimation of the in situ volume, subjectivity of contouring, and patient selection) of the LifeViz 3D system, concluding that the results are comparable with other measurement techniques. The prospects of this technology seem promising in numerous applications in clinical practice to limit the subjectivity of breast surgery.

Entities:  

Mesh:

Year:  2014        PMID: 24511536      PMCID: PMC3910661          DOI: 10.1155/2014/573249

Source DB:  PubMed          Journal:  Biomed Res Int            Impact factor:   3.411


1. Introduction

Breast reconstructive and cosmetic surgery nowadays continues to still be a subjective area where each intervention is surgeon dependent. Currently we note the unfortunate absence of measurement tools that allow objective analysis of this anatomical region, a region that is a particularly representative model of the human variability that limits its study (great variety of shapes, sizes, and compositions). Different volumetric methods of breast measurement (anthropometric formula, Grossman-Roudner method, etc.) have been described in the literature using clinical methods [1-3], casts [4], or the Archimedes principle [5, 6]. The use of medical imaging methods such as mammography [7], sonography [8], and magnetic resonance imagine (MRI) [9] has also been described. However, none of those methods have been reported to be superior to the others. More recently a French group [10] studied a method of optical scanning using structured light projection (Inspeck system) allowing them to assess the advantages and limitations of 3D imaging in breast surgery and concluding that such a volumetric calculation is perfectly suited to clinical practice. Other groups were interested in comparing [11] different methods of volumetry (mammography, the Grossman-Roudner method, anthropometric calculation, the Archimedes principle, and casts) and demonstrated the superiority of imaging, followed by the Archimedes principle. Lately, innovative studies [12] suggested the basis for a new biochemical imaging technology (nanodiamond imaging) that noninvasively records the distribution in two or three dimensions of biologically labeled nanodiamonds in vivo. Our study aims to evaluate and compare (using CT-scanning and the Archimedes principle) a new technique of 3D scanning acquired via stereo-visual technology (3D LifeViz) and its application to biometrics of the breast in clinical practice in terms of volume calculation and sternal notch to nipple distance. Our study included several steps. Experimental section, consisting of 3 components: comparison of three volumetric techniques on anatomical prostheses with known volumes (control); comparison of three volumetric techniques on “anatomical models” (dissection and sampling from cadavers); evaluation of the LifeViz 3D camera by in situ volumetric analysis and on dissections from cadavers. Clinical section, with in situ patient acquisition during consultation for cosmetic and plastic surgery for the assessment of the benefits and limitations of this 3D scanning technique.

2. Materials and Methods

2.1. LifeViz 3D System (Quantificare SA)

LifeViz (Quantificare SA, 1180 Route des Dolines, Athena B, BP 40051, 06901 Sophia Antipolis, France) is a system for 3D reconstruction and analysis (Figure 1).
Figure 1

LifeViz 3D Camera.

LifeViz is composed of a digital stereo camera, a repositioning device (we used a portable system), and 3D software: LifeViz DermaPix 3D from Quantificare and its quantification platform, LifeViz. LifeViz has two main features for volume measurements: LifeViz “Absolute” for the evaluation of the volume in a single acquisition; LifeViz “Evolution” for the relative quantification of volume variation. The principle of LifeViz “Absolute” is the measurement of a cavity or a protrusion using simple image acquisition. It is based on the delineation of the cavity (or protrusion) and then closing the volume with a “minimum surface,” which is the mathematical equivalent of a soap film stretched across the delineated outline. LifeViz “Evolution” is ideal for evaluation of changes in volume over time (it is well adapted to measure the boundaries of anatomical structures that are difficult to define as in the case of labial region).

2.2. Anatomical Specimens

We used nine cadavers of women aged over 60 from the Institute of Human Anatomy (CHU Liège) to measure the volume of each breast in situ and then after dissection (removal of the anatomical section). We used four “prepared” cadavers that had undergone preparation with zinc chloride and preservatives such as ammonium. These bodies were kept in containers with the above-mentioned products for a period of 12 to 18 months (the cadavers were assigned to the Institute of Human Anatomy for medical student teaching). Five other cadavers were “fresh” cadavers frozen at −20°C in the cold room without any preparation. We allowed them to thaw at room temperature in the Institute of Human Anatomy (15°C) for 48 hours before acquisition and dissection. These cadavers had been frozen for more than 6 months.

2.2.1. In Situ Volume Calculation from the Cadavers

First we performed a contouring that consisted of delineating the contours of the breast as closely as possible, subluxing the breast along different axes, defining the furrows to be taken as an outline, and then redrawing the contours with indelible ink. The bodies were all in the dorsal decubitus position for easy mobilization, with the upper limbs flexed to minimize the hidden parts of the breast in 3D acquisitions (to minimize the possibility of a measurement bias and therefore a truncated volume). We took seven imaging shots in total: five views termed “stitching”: front, profile 45° bilaterally, and inferior profile 45° also bilaterally, with a focal distance of 100 cm; two views termed the “incident breast” to minimize hidden parts: inferior-lateral profile 30° bilaterally, with the same focal distance as before (Figure 2).
Figure 2

Contouring technique.

2.2.2. Volume Calculation on Anatomical “Samples” (Experimental)

The breasts were previously marked (left/right) and then carefully resected (as contoured) avoiding taking muscle fibers from the pectoralis major muscle or the serratus anterior muscle. It should be noted that the greatest difficulty of dissection was in the frozen cadavers, where the physiological cleavage planes were almost absent and in which a more difficult but ultimately satisfactory dissection was performed. The 3D LifeViz acquisition was performed with a single lateral view using the table since images taken perpendicularly could overestimate the density. Anatomical sections were placed on a table covered with textured stainless steel forming the surgical field. It is important to note that the “absolute” measurement with the 3D LifeViz camera is accurate when the closing surface is flat, as is the case in this anatomical resection (i.e., when placed flat on the stainless steel table). We performed the same procedure for our control measurements on prostheses of known volume (Mentor). For the volume calculation in situ the closing surface of the 3D volume is somewhat arbitrary. There is a minimum area in a 3D complex shape based on the 3D contour defined by the operator. Due to the physiological curvature of the chest, as well as the fact that resection will create a depression in the body, it is not expected that the in situ measurement will correspond to the measurement of a removed anatomical part (i.e., underestimation of the volume in situ) (Figures 3, 4, and 5).
Figure 3

Schematic presentation of in situ underestimation.

Figure 4

Reconstruction of a breast in situ with 3D LifeViz. Note the shape of the closing surface (white) below. This is a minimum area based on the 3D (black line) contour that is generally concave due to the physiological curvature of the thoracic cage.

Figure 5

Reconstruction of resected breast with 3D LifeViz (anatomical part). The base is a table surface (map), where the 3D contour is quite flat, as this is also the case for the closing surface (white).

2.2.3. In Situ Volume Calculation in Patient

We also undertook 3D image acquisition of 38 patients that presented for various breast-related indications at the Department of Aesthetic and Reconstructive Surgery. These patients were seen preoperatively and, if necessary, during irregular postoperative intervals. Patients were informed of the experimental nature of procedures. The image acquisition procedure comprising seven views was the same as that used in the cadavers, specifically 5 “stitching” views and 2 “incident breast” views. Patients stood with their back against the wall and with the posterior parts of the shoulders abutting the wall to reveal the chest in its entirety. The elbows were bent and hands were placed on the hips to minimize to the greatest extent any hidden part of the breast. Overall, this position gives a good overview of the region of interest in order to observe both breasts and their proper insertions. The position was not very comfortable for the patient, but the speed of acquisition of less than 1 minute made it tolerable. Unable to achieve contouring in our patients during the consultation, we used marking with circular adhesive pads (which were easy to use and safe and had low cost) that were applied in relation to prominent parts of the bones (jugular notch of manubrium sterni (called the “inlet”), the xiphoid appendix, and the acromioclavicular joints bilaterally) and thus delimited the anatomical region of interest (see Figure 6). On the breast itself, the challenge was to define its upper insertion. For this we used a technique of superior subluxation according to Professor P. Blondeel at the University of Ghent (Belgium). The patient raised her breast at its base with the dorsal surface of the ipsilateral hand delimiting an upper portion taken as proximal part of insertion of the breast that we marked with our adhesive pads (Figure 6).
Figure 6

Anatomical spotting technique.

2.3. Archimedes' Principle

We used an old method of Archimedes' principle of buoyancy telling; “Any object, when wholly or partially immersed in a fluid, is buoyed by a force equal to the weight of the fluid displaced by the object” (the volume is equal to the immersed volume of the body). This method is attractive because of its simplicity, reproducibility, low cost, and speed of execution. For this we used a Brand measuring cup of 3000 mL/50 mL by graduation. Then each freshly collected anatomical resected breast was immersed in a determined volume of liquid, in this case 2000 mL of water (the volume was the same for all measures). After recording the different displaced fluid values, we satisfactorily calculated the various results. We repeated the maneuver for the 18 breasts (cadaveric) as well as for four anatomic prostheses of known volume.

2.4. CT-Scan

For scan acquisitions we used a GE Bright Speed 16 strips, model 2010. For this purpose we used the volumetric software “Paint on Slice,” an application of segmentation program “Advantage Work Station GE 4.6” (basic volume viewer). We undertook volume acquisition in spiral mode (abdomen program) without injection of contrast, in axial sections with cuts of 1.25 mm (thickness) every 1 mm (interval), and with 120 kV for 209 mAs per procedure. The acquisition time was 20 seconds for the first series (anatomical parts) and 13.83 sec for the second series (prostheses).

3. Results

3.1. Data Handling and Statistics

We measured the volume (cm³) of the left and right breasts of nine cadavers (S1–S9) using our various techniques. The VIZ3D technique used a 3D camera for volume calculation. This technique was performed once in situ (VIZ3D_INSITU) and once on the resected sample (VIZ3D_PREL). The other two techniques were performed only on the resected sample, namely, Archimedes' manual method (ARCHI_PREL) and the CT scan (CTSCAN_PREL). The volume was also measured by the 3 techniques (VIZ3D_PREL, ARCHI_PREL, and CTSCAN_PREL) in 4 breast implants with different sizes termed controls of known volume (VOLCONNU). We performed measurements in triplicate on the cadaver or anatomical part, the prosthesis or in vivo, and calculated the mean volume measurement for each method, rounded up to the closest unit. We tested three types of 3D LifeViz cameras that differed only by the focal distance of acquisition of 70, 80, and 100 cm. After testing, we kept the camera focal length of 100 cm to allow for optimal breast measurement in our study.

3.2. Data Listing (Table 1)

3.2.1. Methodology

To assess concordance between two series of measurements (technique 1 and technique 2) several statistical methods can be used. Coefficient of Intraclass Correlation (ICC). The ICC is a measure of concordance between two series where the studied variable is continuous (the volume measures in cm³). The closer the ICC is to 1, the better the concordance is between the two series. Zero (ICC = 0) signifies the absence of concordance. Paired Samples Student's t-test/Wilcoxon Signed Rank Test. The paired samples Student's t-test is used to test the hypothesis that the average difference between two measures (techniques) is absent. A P value is associated with value obtained from Student's t-test. If the test is rejected (P < 0.05), that means there is a systematic difference between the values provided by two techniques. Otherwise, we consider that the means in two techniques give the same values. Wilcoxon test of signed ranks is a nonparametric test corresponding to the Student's t-test and compares the medians. Reproducibility of Measurements (CV). To measure the reproducibility of a test, namely, the ability to reproduce the same value by repeating the measurement, we can use two results obtained for each sample as if they had both been obtained using the same technique. Reproducibility is best expressed by a coefficient of variation CV (%). The procedure is as follows: consider n pairs of measurements {(x , x ), i = 1,…, N}. Note that and s² = (∑d 2)/(2n), where is the mean of means of two measurements and {d = x − x , i = 1,…, N}. So, . The lower the CV is, the better the reproducibility is. Bland-Altman Plot. In a Bland-Altman plot, for each pair of measurements {(x , x ),  i = 1,…, N} we report on the abscissa the mean of two measurements and on the ordinal the difference d = x − x . In principle, there should not be any association between d and . We can test this hypothesis by calculating the correlation coefficient r (classical or spearman) between the two series and associate it to the P value. The absence of correlation indicates that the gap between two measurements does not vary with volume.

3.3. Cadavers: Comparison of VIZ3D In Situ versus VIZ3D Sampling (Table 2)

There was a significant difference of 41.4 ± 68.9 cm³ (P = 0.021) between two methods. The in situ method underestimates the volume compared to the resection-based method. The concordance between the two methods is good (0.87) but the lower confidence limit is not very high. The difference between two methods is 19.9% (Figure 7).
Figure 7

Bland-Altman plot comparing VIZ3D in situ versus VIZ3D sample from the bodies. This graph shows stability of measures; the difference between the measurements does not depend on volume.

Regarding the underestimation of in situ measurements, additional information could be assessed. The 3D image could be obtained after resection of the anatomical area. This would allow measurement by the 3D LifeViz camera of the hollow anatomical section left by the resection. This would allow one to compare the differences in volume before and after resection, the volume of the resected portion, which would explain the underestimation of the volume due to the hollow left by resection. For its part, the company trained operator also made measurements of the same areas as us. Note that all their measurements were made with a single view of the 3D camera (instead of 5 + 2 made by us), with the angle that seemed the most appropriate. This was possible in all subjects except one case in situ and in one localization for this patient, where the arm limited this measurement. As a result, the company excluded this particular image of the in situ experiment and therefore used 8 subjects instead of 9 in our series. Their number 7 has only one image on the right side because the left side contained too many hidden regions and therefore could bias the measurement. They did not perform the mean of three measurements (rather, one single measurement in the optimal orientation of imaging was used). Their trained experience in shooting and contouring (one large contour will permit better definition of the closing surface without increasing the measured total volume and has less risk of losing a part of the volume of the object) as compared to “naive” operators is predictable. Their findings after statistical integration are shown. Note that the order of their subjects is not identical to ours (Table 3). We observed a significant difference of 47.1 ± 36.5 cm³ (P = 0.0002) between the two methods. The concordance between the two methods is very good (0.92) but the lower confidence limit is not particularly high. The coefficient of variation between two methods is 14.5%. As expected, the company measurements are more precise and better correlated than ours. This indicated that the measurements are highly operator dependent. Therefore for future clinical use by various practitioners standardized training by the company itself is needed to have reproducibility of measurements (particularly important in clinical follow-up or pre-and postoperative comparison).

3.4. Controls: Comparison of Methods (Table 4)

We used as “controls” four anatomic Mentor implants with known volume (T1: 60cc, T2: 150cc, T3: 330cc, and T4: 460cc). Using these controls we compared one with another using our three methods of study and we found good concordances and relatively low coefficients of variation. However, the results should be assessed with caution because there were only four observations (Figure 8).
Figure 8

Comparisons of methods (controls) based on ICC values. The results should be assessed with caution because there were only four observations.

3.5. Cadavers: Comparison of Methods Using Resected Samples

In the cadavers (S1–S9), when comparing the volume with the reference method (here it was the CT scan), there was a mean difference of 47.2 ± 39.0 cm³ for VIZ3D method and a mean difference of 13.1 ± 23.6 cm³ for the Archimedes method. In all cases, the difference was statistically significant. The best concordance with the CT scan was obtained with the Archimedes method and with this the ICC was 0.992 and the lower confidence limit was very high (0.978). For the VIZ3D method, we obtained a concordance of 0.951 (0.629) between the samples. For the VIZ3D method, the ICC was high but the lower confidence limit was below that observed for the Archimedes method. The coefficient of variation for the Archimedes method was also lower (5.29%) as compared to the VIZ3D technique (in situ 29.9% and on samples 13.2%). With these results, we can suggest that imaging is the technique of choice, followed by the Archimedes principle and then the 3D imaging LifeViz. If we look at the overall concordance between three methods, we obtain an ICC of 0.96 (inferior limit = 0.90), suggesting comparability of these three methods (Figures 9, 10, 11, and 12).
Figure 9

Bland-Altman plot comparing VIZ3D sampling versus CT scan sampling (parts of dissection). When comparing the volume with the reference method (here it was the CT scan), there was a mean difference of 47.2 ± 39.0 cm³ for VIZ3D method.

Figure 10

Bland-Altman plot comparing VIZ3D sampling versus Archimedes' method on samples (anatomical dissection). When comparing the volume with the reference method (here it was the CT scan), there was a mean difference of 13.1 ± 23.6 cm³ for the Archimedes method. In all cases, the difference was statistically significant.

Figure 11

Comparison of methods (cadavers) based on ICC values. The best concordance with the CT scan was obtained with the Archimedes method and with this the ICC was 0.992 and the lower confidence limit was very high (0.978). For the VIZ3D method, we obtained a concordance of 0.951 (0.629) between the samples. For the VIZ3D method, the ICC was high but the lower confidence limit was below that observed for the Archimedes method.

Figure 12

Comparison of methods (cadavers + controls) based on ICC values. If we look at the overall concordance between three methods, we obtain an ICC of 0.96 (inferior limit = 0.90), suggesting comparability of these three methods.

3.6. Acquisition of Patients' Images

Over a period of a month we scanned 38 patients, preoperatively in the majority of cases. These scans allowed us to identify issues related to the acquisition of patients in clinical practice (during consultation) and to gain the necessary experience with a view to clinical use. Indications for surgery in patients who received at least one 3D acquisition were varied, such as, breast reconstruction by prosthesis and/or lipomodelling, remodeling of breast reconstruction, breast asymmetry, breast ptosis, or breast reduction.

4. Discussion

This study evaluates the application of 3D image acquisition to breast surgery using a new technique for 3D stereovision, LifeViz. We demonstrated a good correlation between the measurements made using the CT scan and Archimedes' principle buoyancy with anatomical samples and isolated implants and with the 3D camera. However, the volume calculation in situ (cadaveric and clinical) needs to be improved (due to significant underestimation of volume). The advantages and limitations of this new technology should be considered. Advantages. (i) Portable device: this is a major advantage of this system that could increase clinical use (the transportation to consultation from one department to another), usage in operating rooms (acquisition could be performed pre-, per, and immediately postoperatively or later), mobility (low weight and acceptable size), and usage in any place (no need of a “special” room). We would emphasize the substantial autonomy for several days of use. (ii) Safety: no irradiation, a flash similar to those used in photographic cameras and harmless to the human eye, pain-free technique. (iii) Low stress for the patient: very short acquisition time (of the order of seconds), which increases the quality of measurements by avoiding patient movement (that can produce image distortion). The acquisition can be performed sitting or standing. Here we chose the standing position with hands on hips (intermediate position between the anatomical position and the raised arms position) to minimize the hidden parts of the chest and to maintain the advantages of both positions above (this was possible because our patients were all preoperative); the intermediate position was originally described by Sinna et al. [10]. (iv) Additional value: the advantages over traditional photography (the objective forms angles of view not accessible by standard simple photography). (v) Cost: <15000 Euros for the entire apparatus. (vi) Noninvasive technique: noncontact nature of the process (therefore possible to use even on sensitive soft tissue, and it doesn't cause any deformation). Disadvantages. (i) A problem is that the focal length needs to be respected (between 80 and 120 cm), with the risk of image distortion as a penalty (computer contouring difficulties occur in these conditions) and the problem of comparability of results. (ii) Difficulty in defining breast contours (upper limit is the axillary pillar). (iii) Underestimation of the volume in situ, hence the difficulty of reliable clinical utilization. (iv) Selection of patients: patients with significant ptosis or who are significantly overweight disrupt the scans, resulting in the measurement bias by increasing the hidden regions (in particular that of segment III) but also making the breast limits barely perceptible during contouring. (v) Subjectivity in contouring: contouring is very operator dependent, which makes it difficult to reproduce.

5. Conclusion

This technology appears to offer a promising future because of its multiple applications particularly in clinical practice. The technology challenges the subjectivity of surgery, allowing the more likely obtainment of predictable and defined results, improving patient satisfaction and serving as an objective and reliable measurement tool for the practitioner to improve the quality of interventions and outcomes. In our evaluation of the 3D LifeViz camera, in situ volume calculations alone remain perfect for true routine clinical use.
Table 1

Data listing.

GroupNameSideVIZ3D_INSITUVIZ3D_PRELARCHI_PRELCTSCAN_PRELKnown vol
PatientsS1G205190220219
S1D200180200214
S2L485490550538
S2R380385450433
S3L280315450451
S3R320390480507
S4L70130150146
S4R180220250229
S5L330385425403
S5R285325400364
S6L105140180185
S6R135130170145
S7L125145200191
S7R170175250208
S8L470540600612
S8R5108051000923
S9L205230260244
S9R200225250237

ControlsT175556060
T2157150146150
T3339350335330
T4461450474460

(a)

NameSideVIZ3D_INSITUVIZ3D_SAMPdVIZisVISp
S1G20519015
S1D20018020
S2L485490−5
S2R380385−5
S3L280315−35
S3R320390−70
S4L70130−60
S4R180220−40
S5L330385−55
S5R285325−40
S6L105140−35
S6R1351305
S7L125145−20
S7R170175−5
S8L470540−70
S8R510805−295
S9L205230−25
S9R200225−25

Mean ± SD258.6 ± 133.5300.0 ± 178.0−41.4 ± 68.9
Median−30
P25–P75−55–−5
SD (robust)37.0

(b)

Method 1Method 2 N ICCDifference Paired Student's t-testWilcoxonCV (%)Bland-Altman
(ICC*) P value P value r (P value)
VIZ3D in situVIZ3D sample180.878 (0.698)−41.4 ± 68.90.0210.000819.9Pearson: −0.66 (0.0031)
Spearman: −0.39 (0.10)

There was a significant difference of 41.4 ± 68.9 cm³  (P = 0.021) between two methods. The in situ method underestimates the volume compared to the resection-based method.

(a)

SubjectSideVolume_in_situVolume_resectiondiff
P001Right384.5424.7−40.2
P001Left507.7520.8−13.1
P002Right362.3436.8−74.5
P002Left276.0402.3−126.3
P003Right171.5234.0−62.5
P003Left69.1156.7−87.6
P004Right285.8350.1−64.3
P004Left345.1398.5−53.4
P005Right153.2133.220.0
P005Left108.3173.0−64.7
P006Right168.9205.0−36.1
P006Left123.8165.4−41.6
P007Right525.7574.9−49.2
P008Right244.7248.2−3.5
P008Left230.3240.3−10.0

Mean ± SD264 ± 140311 ± 142−47.1 ± 36.5
Median245248−49.2
P25–P75153–362173–425−64.7–−13.1
SD (robust)15518638.2

(b)

Method 1Method 2 N ICCDifferencePaired Student's t-testWilcoxonCV (%)Bland-Altman
(ICC*) P value P value r (P value)
In situResection150.92 (0.42)−47.1 ± 36.50.00020.000414.5Pearson: r = −0.06  (P = 0.82)
Spearman: r = 0.04  (P = 0.89)

We observed a significant difference 47.1 ± 36.5 cm3  (P = 0.0002) between the two methods. The concordance between the two methods is very good (0.92) but the lower confidence limit is not particularly high.

Table 4

Comparison of methods (controls).

Method 1Method 2 N ICCDifferenceStudent's t-test WilcoxonCV (%)Bland-Altman
(ICC*) P value P value r (P value)
T1–T4
 Known volumeVIZ3D samp.40.999 (0.968)−8.00 ± 5.770.0700.132.63Pearson: 0.84 (0.16)
Spearman: 0.80 (0.20)
 Known volumeArchimedes' samp.40.999 (0.982)−1.25 ± 13.10.860.993.23Pearson: −0.10 (0.90)
Spearman: 0.20 (0.80)
 Known volumeCT scan samp.40.999 (0.994)−3.75 ± 7.760.410.502.16Pearson: −0.899 (0.10)
Spearman: −0.80 (0.20)
 VIZ3D samp.Archimedes' samp.40.997 (0.982)6.75 ± 13.00.380.503.65Pearson:−0.47 (0.53)
Spearman: −0.40 (0.60)
 VIZ3D samp.CT scan samp40.998 (0.986)4.25 ± 12.40.540.633.18Pearson: −0.956 (0.044)
Spearman: −1.00 (<0.0001)
 Archimedes' samp.CT scan samp40.997 (0.973)−2.50 ± 16.50.780.884.06Pearson: −0.34 (0.66)
Spearman: −0.20 (0.80)

Using these controls we compared one with another using our three methods of study and we found good concordances and relatively low coefficients of variation.

  12 in total

1.  Mammographic determination of breast volume: comparing different methods.

Authors:  C L Kalbhen; J J McGill; P M Fendley; K W Corrigan; J Angelats
Journal:  AJR Am J Roentgenol       Date:  1999-12       Impact factor: 3.959

2.  [Evaluation of 3D numerisation with structured light projection in breast surgery].

Authors:  R Sinna; S Garson; F Taha; T Benhaim; C Carton; E Delay; M Robbe
Journal:  Ann Chir Plast Esthet       Date:  2009-05-06       Impact factor: 0.660

3.  Measurement of breast volume with thermoplastic casts.

Authors:  A Edsander-Nord; M Wickman; G Jurell
Journal:  Scand J Plast Reconstr Surg Hand Surg       Date:  1996-06

4.  Nanodiamond imaging: a new molecular imaging approach.

Authors:  Alex N Hegyi; Eli Yablonovitch
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2012

5.  Breast volume measurement in young Chinese women and clinical applications.

Authors:  Q Qiao; G Zhou; Y Ling
Journal:  Aesthetic Plast Surg       Date:  1997 Sep-Oct       Impact factor: 2.326

6.  Further applications of Archimedes' principle in the correction of asymmetrical breasts.

Authors:  R C Schultz; R F Dolezal; J Nolan
Journal:  Ann Plast Surg       Date:  1986-02       Impact factor: 1.539

7.  Volumetric measurement of the human breast and breast tissue before and during mammaplasty.

Authors:  F G Bouman
Journal:  Br J Plast Surg       Date:  1970-07

8.  Measurement of breast implant volume with magnetic resonance imaging.

Authors:  M Mineyev; D Kramer; L Kaufman; J Carlson; S Frankel
Journal:  Ann Plast Surg       Date:  1995-04       Impact factor: 1.539

9.  Volumetric estimation in breast surgery.

Authors:  A M Morris
Journal:  Br J Plast Surg       Date:  1978-01

10.  Five methods of breast volume measurement: a comparative study of measurements of specimen volume in 30 mastectomy cases.

Authors:  Ragip Kayar; Serdar Civelek; Murat Cobanoglu; Osman Gungor; Hidayet Catal; Mustafa Emiroglu
Journal:  Breast Cancer (Auckl)       Date:  2011-03-27
View more
  10 in total

1.  Contralateral breast volume measurement during chest CT for postmastectomy breast reconstruction.

Authors:  Noha Mohamed Osman; Samer Malak Botros; Ahmed Fathy Abdel Ghany; Ashraf Maher Farid
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-05-08       Impact factor: 2.924

2.  Feasibility of an electrostimulation system treatment for wound healing: a case series of patients with chronic ulcers in Barbados.

Authors:  Nkemcho Ojeh; Angela Rose; Selma Jackman; Morvillia Applewhaite; Veronica Webster
Journal:  Int Wound J       Date:  2015-03-10       Impact factor: 3.315

3.  A New Protocol to Treat Abdominal Subcutaneous Fat Combining Microwaves and Flat magnetic stimulation.

Authors:  Steven Paul Nisticò; Paolo Bonan; Federica Coli; Alice Verdelli; Irene Fusco; Francesco Gratteri; Claudia Sicilia; Carmen Cantisani; Giovanni Pellacani; Luigi Bennardo; Giovanni Cannarozzo
Journal:  Bioengineering (Basel)       Date:  2022-04-21

4.  Breast Anthropometry-Results of a Prospective Study Among Indian Breast Cancer Patients.

Authors:  Praveen Royal Mokkapati; Manoj Gowda; Suryanarayana Deo; Ekta Dhamija; Sanjay Thulkar
Journal:  Indian J Surg Oncol       Date:  2020-01-07

5.  The potential role of three-dimensional surface imaging as a tool to evaluate aesthetic outcome after Breast Conserving Therapy (BCT).

Authors:  Rachel L O'Connell; Rosa Di Micco; Komel Khabra; Lisa Wolf; Nandita deSouza; Nicola Roche; Peter A Barry; Anna M Kirby; Jennifer E Rusby
Journal:  Breast Cancer Res Treat       Date:  2017-04-26       Impact factor: 4.872

6.  New software and breast boundary landmarks to calculate breast volumes from 3D surface images.

Authors:  T S Wesselius; R D Vreeken; A C Verhulst; T Xi; T J J Maal; D J O Ulrich
Journal:  Eur J Plast Surg       Date:  2018-07-06

7.  Stereophotogrammetric three-dimensional photography is an accurate and precise planimetric method for the clinical visualization and quantification of human papilloma virus-induced skin lesions.

Authors:  M Rijsbergen; L Pagan; T Niemeyer-van der Kolk; R Rijneveld; G Hogendoorn; C Lemoine; Y Meija Miranda; G Feiss; J N Bouwes Bavink; J Burggraaf; M I E van Poelgeest; R Rissmann
Journal:  J Eur Acad Dermatol Venereol       Date:  2019-03-12       Impact factor: 6.166

8.  Patient-Reported Measurement of Breast Asymmetry Using Archimedes' Principle in Breast Reduction Mammaplasty: A Retrospective Study.

Authors:  Daniel Waltho; Mark McRae; Achilleas Thoma
Journal:  Cureus       Date:  2020-01-01

Review 9.  A systematic review of objective burn scar measurements.

Authors:  Kwang Chear Lee; Janine Dretzke; Liam Grover; Ann Logan; Naiem Moiemen
Journal:  Burns Trauma       Date:  2016-04-27

10.  "Three-dimensional evaluation of breast volume changes following autologous free flap breast reconstruction over six months".

Authors:  Floor N H Wilting; Marijn Hameeteman; Hanneke J P Tielemans; Dietmar J O Ulrich; Stefan Hummelink
Journal:  Breast       Date:  2020-02-10       Impact factor: 4.380

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.