Literature DB >> 24510421

Retrograde trans-anterior communicating artery rescue of unopened Pipeline Embolization Device with balloon dilation: complication management.

Ramon Navarro1, Jang Yoon2, Tanya Dixon1, David A Miller3, Ricardo A Hanel2, Rabih G Tawk2.   

Abstract

As the use of the Pipeline Embolization Device (PED) for the treatment of complex intracranial aneurysms rises, knowledge about complications continues to accumulate amidt a paucity of reports on techniques and rescue strategies. We describe the case of a 70-year-old woman who presented with worsening reto-orbital left-sided pain and a large cavernous aneurysm. The patient underwent endovascular treatment with PED, and there was difficulty delivering the device due to significant vascular tortuosity. This resulted in poor PED deployment as the proximal end failed to open. Increasingly aggressive strategies were attempted to open the device, which resulted in an iatrogenic carotid cavernous fistula. We were finally able to rescue the device and open its proximal end with balloon inflation after using a contralateral trans-anterior communicating artery approach and crossing the PED in a retrograde fashion. Excessive vascular tortuosity poses a genuine risk of PED malfunction and poor deployment. Although we were able to rescue the device and our patient had no permanent morbidity, difficult vascular anatomy rendered the procedure extremely complicated with dreaded complications. Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://group.bmj.com/group/rights-licensing/permissions.

Entities:  

Keywords:  Aneurysm; Complication; Device; Fistula

Mesh:

Year:  2014        PMID: 24510421     DOI: 10.1136/neurintsurg-2013-011009.rep

Source DB:  PubMed          Journal:  J Neurointerv Surg        ISSN: 1759-8478            Impact factor:   5.836


  3 in total

1.  Flow Diversion for Intracranial Aneurysm Management: A New Standard of Care.

Authors:  Fawaz Al-Mufti; Krishna Amuluru; Chirag D Gandhi; Charles J Prestigiacomo
Journal:  Neurotherapeutics       Date:  2016-07       Impact factor: 7.620

Review 2.  Flow diverters for intracranial aneurysms.

Authors:  Yazan J Alderazi; Darshan Shastri; Tareq Kass-Hout; Charles J Prestigiacomo; Chirag D Gandhi
Journal:  Stroke Res Treat       Date:  2014-05-20

3.  Inadvertent Complication of a Pipeline Embolization Device for Treatment with Vertebral Artery Dissecting Aneurysm : Distal Tip Fracture of Delivery Wire.

Authors:  Jung Soo Park; Hyo Sung Kwak; Jong Myong Lee
Journal:  J Korean Neurosurg Soc       Date:  2016-09-08
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.