Literature DB >> 2451036

Differences in values obtained with 2 assays of prostate specific antigen.

G L Hortin1, R R Bahnson, M Daft, K M Chan, W J Catalona, J H Ladenson.   

Abstract

Several different assays for prostate specific antigen have become available for use in clinical laboratories. We compared the 2 most widely used assays, the Tandem-R and the Pros-Check prostate specific antigen assays, to determine if different assays yield comparable results. We analyzed 70 serum specimens from patients with suspected or known prostatic carcinoma by both assays. Results from the assays showed close linear correlation (r equals 0.988) but the Pros-Check assay yielded values 1.85 times those of the Tandem-R test. The proportional bias between assays is owing to differences in the values assigned to calibration standards in the assays, and it demonstrates a need for improved standardization of prostate specific antigen assays. Important consequences of the bias between assays are that different normal ranges apply and that values from the 2 assays cannot be compared directly.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 2451036     DOI: 10.1016/s0022-5347(17)42627-9

Source DB:  PubMed          Journal:  J Urol        ISSN: 0022-5347            Impact factor:   7.450


  2 in total

1.  Development of an immunoassay specific for the PSA-ACT complex without the problem of high background.

Authors:  J T Wu; P Zhang; G H Liu; L Wilson
Journal:  J Clin Lab Anal       Date:  1998       Impact factor: 2.352

2.  Molecular Form Differences Between Prostate-Specific Antigen (PSA) Standards Create Quantitative Discordances in PSA ELISA Measurements.

Authors:  Erica L McJimpsey
Journal:  Sci Rep       Date:  2016-02-25       Impact factor: 4.379

  2 in total