Dependency of EBT2 film calibration curve on postirradiation time.

Liyun Chang1, Sheng-Yow Ho2, Hueisch-Jy Ding1, Tsair-Fwu Lee3, Pang-Yu Chen4.   

Abstract

PURPOSE: The Ashland Inc. product EBT2 film model is a widely used quality assurance tool, especially for verification of 2-dimensional dose distributions. In general, the calibration film and the dose measurement film are irradiated, scanned, and calibrated at the same postirradiation time (PIT), 1-2 days after the films are irradiated. However, for a busy clinic or in some special situations, the PIT for the dose measurement film may be different from that of the calibration film. In this case, the measured dose will be incorrect. This paper proposed a film calibration method that includes the effect of PIT.
METHODS: The dose versus film optical density was fitted to a power function with three parameters. One of these parameters was PIT dependent, while the other two were found to be almost constant with a standard deviation of the mean less than 4%. The PIT-dependent parameter was fitted to another power function of PIT. The EBT2 film model was calibrated using the PDD method with 14 different PITs ranging from 1 h to 2 months. Ten of the fourteen PITs were used for finding the fitting parameters, and the other four were used for testing the model.
RESULTS: The verification test shows that the differences between the delivered doses and the film doses calculated with this modeling were mainly within 2% for delivered doses above 60 cGy, and the total uncertainties were generally under 5%. The errors and total uncertainties of film dose calculation were independent of the PIT using the proposed calibration procedure. However, the fitting uncertainty increased with decreasing dose or PIT, but stayed below 1.3% for this study.
CONCLUSIONS: The EBT2 film dose can be modeled as a function of PIT. For the ease of routine calibration, five PITs were suggested to be used. It is recommended that two PITs be located in the fast developing period (1 ∼ 6 h), one in 1 ∼ 2 days, one around a week, and one around a month.

Mesh:

Year:  2014        PMID: 24506617     DOI: 10.1118/1.4862511

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Evaluation of a single-scan protocol for radiochromic film dosimetry.

Authors:  Yoshinobu Shimohigashi; Fujio Araki; Masato Maruyama; Yuji Nakaguchi; Satoshi Kuwahara; Nozomu Nagasue; Yudai Kai
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

2.  Calibration of Gafchromic EBT Film Using the Microtek ScanMaker 9800XL Plus Flatbed Scanner with a Modified One Red-Channel after Three-Channel Method.

Authors:  Liyun Chang; Sheng-Yow Ho; Hueisch-Jy Ding; Shyh-An Yeh; Pang-Yu Chen
Journal:  J Med Phys       Date:  2019 Jul-Sep

3.  Calibration of the EBT3 Gafchromic Film Using HNN Deep Learning.

Authors:  Liyun Chang; Shyh-An Yeh; Sheng-Yow Ho; Hueisch-Jy Ding; Pang-Yu Chen; Tsair-Fwu Lee
Journal:  Biomed Res Int       Date:  2021-01-31       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.