Development and evaluation of a prototype tracking system using the treatment couch.

Stephanie Lang1, Jörg Zeimetz2, Gregor Ochsner3, Marianne Schmid Daners3, Oliver Riesterer1, Stephan Klöck1.   

Abstract

PURPOSE: Tumor motion increases safety margins around the clinical target volume and leads to an increased dose to the surrounding healthy tissue. The authors have developed and evaluated a one-dimensional treatment couch tracking system to counter steer respiratory tumor motion. Three different motion detection sensors with different lag times were evaluated.
METHODS: The couch tracking system consists of a motion detection sensor, which can be the topometrical system Topos (Cyber Technologies, Germany), the respiratory gating system RPM (Varian Medical Systems) or a laser triangulation system (Micro Epsilon), and the Protura treatment couch (Civco Medical Systems). The control of the treatment couch was implemented in the block diagram environment Simulink (MathWorks). To achieve real time performance, the Simulink models were executed on a real time engine, provided by Real-Time Windows Target (MathWorks). A proportional-integral control system was implemented. The lag time of the couch tracking system using the three different motion detection sensors was measured. The geometrical accuracy of the system was evaluated by measuring the mean absolute deviation from the reference (static position) during motion tracking. This deviation was compared to the mean absolute deviation without tracking and a reduction factor was defined. A hexapod system was moving according to seven respiration patterns previously acquired with the RPM system as well as according to a sin(6) function with two different frequencies (0.33 and 0.17 Hz) and the treatment table compensated the motion.
RESULTS: A prototype system for treatment couch tracking of respiratory motion was developed. The laser based tracking system with a small lag time of 57 ms reduced the residual motion by a factor of 11.9 ± 5.5 (mean value ± standard deviation). An increase in delay time from 57 to 130 ms (RPM based system) resulted in a reduction by a factor of 4.7 ± 2.6. The Topos based tracking system with the largest lag time of 300 ms achieved a mean reduction by a factor of 3.4 ± 2.3. The increase in the penumbra of a profile (1 × 1 cm(2)) for a motion of 6 mm was 1.4 mm. With tracking applied there was no increase in the penumbra.
CONCLUSIONS: Couch tracking with the Protura treatment couch is achievable. To reliably track all possible respiration patterns without prediction filters a short lag time below 100 ms is needed. More scientific work is necessary to extend our prototype to tracking of internal motion.

Entities:  

Mesh:

Year:  2014        PMID: 24506611     DOI: 10.1118/1.4862077

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  Technical Note: In silico and experimental evaluation of two leaf-fitting algorithms for MLC tracking based on exposure error and plan complexity.

Authors:  Vincent Caillet; Ricky O'Brien; Douglas Moore; Per Poulsen; Tobias Pommer; Emma Colvill; Amit Sawant; Jeremy Booth; Paul Keall
Journal:  Med Phys       Date:  2019-03-04       Impact factor: 4.071

2.  Robotic ultrasound-guided SBRT of the prostate: feasibility with respect to plan quality.

Authors:  Stefan Gerlach; Ivo Kuhlemann; Philipp Jauer; Ralf Bruder; Floris Ernst; Christoph Fürweger; Alexander Schlaefer
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-07-12       Impact factor: 2.924

3.  A dosimetric comparison of real-time adaptive and non-adaptive radiotherapy: A multi-institutional study encompassing robotic, gimbaled, multileaf collimator and couch tracking.

Authors:  Emma Colvill; Jeremy Booth; Simeon Nill; Martin Fast; James Bedford; Uwe Oelfke; Mitsuhiro Nakamura; Per Poulsen; Esben Worm; Rune Hansen; Thomas Ravkilde; Jonas Scherman Rydhög; Tobias Pommer; Per Munck Af Rosenschold; Stephanie Lang; Matthias Guckenberger; Christian Groh; Christian Herrmann; Dirk Verellen; Kenneth Poels; Lei Wang; Michael Hadsell; Thilo Sothmann; Oliver Blanck; Paul Keall
Journal:  Radiother Oncol       Date:  2016-03-22       Impact factor: 6.280

4.  Unconscious physiological response of healthy volunteers to dynamic respiration-synchronized couch motion.

Authors:  Alexander Jöhl; Marta Bogowicz; Stefanie Ehrbar; Matthias Guckenberger; Stephan Klöck; Mirko Meboldt; Oliver Riesterer; Melanie Zeilinger; Marianne Schmid Daners; Stephanie Tanadini-Lang
Journal:  Radiat Oncol       Date:  2017-11-28       Impact factor: 3.481

5.  Long-term stability assessment of a 4D tumor tracking system integrated into a gimbaled linear accelerator.

Authors:  Mami Akimoto; Mitsuhiro Nakamura; Yuki Miyabe; Nobutaka Mukumoto; Kenji Yokota; Takashi Mizowaki; Masahiro Hiraoka
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

6.  Potential dosimetric benefits of adaptive tumor tracking over the internal target volume concept for stereotactic body radiation therapy of pancreatic cancer.

Authors:  Konstantina Karava; Stefanie Ehrbar; Oliver Riesterer; Johannes Roesch; Stefan Glatz; Stephan Klöck; Matthias Guckenberger; Stephanie Tanadini-Lang
Journal:  Radiat Oncol       Date:  2017-11-09       Impact factor: 3.481

7.  Optimization based trajectory planning for real-time 6DoF robotic patient motion compensation systems.

Authors:  Xinmin Liu; Rodney D Wiersma
Journal:  PLoS One       Date:  2019-01-11       Impact factor: 3.240

8.  The potential of Gantry beamline large momentum acceptance for real time tumour tracking in pencil beam scanning proton therapy.

Authors:  Giovanni Fattori; Ye Zhang; David Meer; Damien Charles Weber; Antony John Lomax; Sairos Safai
Journal:  Sci Rep       Date:  2020-09-18       Impact factor: 4.379

9.  Direct measurement and correction of both megavoltage and kilovoltage scattered x-rays for orthogonal kilovoltage imaging subsystems with dual flat panel detectors.

Authors:  Hiraku Iramina; Mitsuhiro Nakamura; Takashi Mizowaki
Journal:  J Appl Clin Med Phys       Date:  2020-07-25       Impact factor: 2.102

10.  Fast Fourier transform combined with phase leading compensator for respiratory motion compensation system.

Authors:  Chia-Chun Kuo; Ho-Chiao Chuang; Ai-Ho Liao; Hsiao-Wei Yu; Syue-Ru Cai; Der-Chi Tien; Shiu-Chen Jeng; Jeng-Fong Chiou
Journal:  Quant Imaging Med Surg       Date:  2020-05
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.