Literature DB >> 24500941

PIKO-6® vs. forced spirometry in asthmatic children.

Laura Gochicoa-Rangel1, Pablo José Larios-Castañeda, José Luis Miguel-Reyes, David Martínez Briseño, Roberto Flores-Campos, Juan Arturo Sáenz-López, Luis Torre-Bouscoulet.   

Abstract

BACKGROUND: The PIKO-6® is an electronic device that measures forced expiratory volume at seconds 1 (FEV1) and 6 (FEV6) of a forced vital capacity (FVC) maneuver. This device could aid in diagnosing obstructive respiratory diseases.
OBJECTIVES: To determine the concordance of FEV1, FEV6, and the FEV1/FEV6 quotient achieved with PIKO-6® versus spirometric values from asthmatic patients, and compare results with measures from healthy children.
METHODS: A cross-sectional study with asthmatic and healthy 6-to-14-year-old children, all of whom performed a forced spirometry as well as a PIKO-6® test.
RESULTS: The study included 82 subjects (58 asthmatics, 24 healthy children). Except for the functional parameters, the basal characteristics of the two groups were similar. The concordance correlation coefficient (CCC) for FEV1 was 0.938 (P < 0.001), with 95% limits of agreement of -0.591 to 0.512 L, and an average of differences of -0.040 L. For FEV6, CCC was 0.927 (P < 0.001), and the 95% limits of agreement were -0.751 to 0.598 L with an average of differences of -0.077 L. The concordance analysis and the FEV1 and FEV6 associations were better in children with controlled asthma and healthy subjects, as well as in the post-bronchodilator results.
CONCLUSIONS: The concordance between PIKO-6® and spirometry was lower in patients with partially controlled or uncontrolled asthma compared to controlled or healthy children. The broad limits of agreement show that the FEV1, FEV6, and FEV1/FEV6 obtained with the PIKO-6® are not interchangeable with spirometry results. Longitudinal evaluations of asthma patients are necessary to assess the utility of PIKO-6®.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  FEV1; FEV6; FVC; PIKO-6®; asthmatic children; peak flow; spirometry

Mesh:

Year:  2014        PMID: 24500941     DOI: 10.1002/ppul.22996

Source DB:  PubMed          Journal:  Pediatr Pulmonol        ISSN: 1099-0496


  3 in total

1.  Utility of a portable hand-held forced expiratory volume in 1 second meter for monitoring asthma in an inner-city school setting.

Authors:  Jonathan M Gaffin; Joseph S Zhou; Qiling Cai; Carter R Petty; Chunxia Fu; William J Sheehan; Sachin Baxi; Ann Bailey; Diane R Gold; Wanda Phipatanakul
Journal:  Ann Allergy Asthma Immunol       Date:  2016-07-14       Impact factor: 6.347

2.  Bisphenol a exposure and the development of wheeze and lung function in children through age 5 years.

Authors:  Adam J Spanier; Robert S Kahn; Allen R Kunselman; Eric W Schaefer; Richard Hornung; Yingying Xu; Antonia M Calafat; Bruce P Lanphear
Journal:  JAMA Pediatr       Date:  2014-12       Impact factor: 16.193

3.  Peak flow meter with a questionnaire and mini-spirometer to help detect asthma and COPD in real-life clinical practice: a cross-sectional study.

Authors:  Yogesh T Thorat; Sundeep S Salvi; Rahul R Kodgule
Journal:  NPJ Prim Care Respir Med       Date:  2017-05-09       Impact factor: 2.871

  3 in total