Literature DB >> 24497287

Matrix-assisted laser desorption/ionization time-of-flight mass spectrometric imaging of synthetic polymer sample spots prepared using ionic liquid matrices.

Stefan J Gabriel1, Dietmar Pfeifer, Clemens Schwarzinger, Ulrich Panne, Steffen M Weidner.   

Abstract

RATIONALE: Polymer sample spots for matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) prepared by the dried-droplet method often reveal ring formation accompanied by possible segregation of matrix and sample molecules as well as of the polymer homologs itself. Since the majority of sample spots are prepared by this simple and fast method, a matrix or sample preparation method that excludes such segregation has to be found.
METHODS: Three different ionic liquid matrices based on conventionally used aromatic compounds for MALDI-TOF MS were prepared. The formation of ionic liquids was proven by (1) H NMR spectroscopy. MALDI-Imaging mass spectrometry was applied to monitor the homogeneity.
RESULTS: Our results show a superior sample spot homogeneity using ionic liquid matrices. Spots could be sampled several times without visible differences in the mass spectra. A frequently observed loss of matrix in the mass spectrometer vacuum was not observed. The necessary laser irradiance was reduced, which resulted in less polymer fragmentation.
CONCLUSIONS: Ionic liquid matrices can be used to overcome segregation, a typical drawback of conventional MALDI dried-droplet preparations. Homogeneous sample spots are easy to prepare, stable in the MS vacuum and, thereby, improve the reproducibility of MALDI.
Copyright © 2014 John Wiley & Sons, Ltd.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24497287     DOI: 10.1002/rcm.6810

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  2 in total

1.  Reducing Spatial Heterogeneity of MALDI Samples with Marangoni Flows During Sample Preparation.

Authors:  Yin-Hung Lai; Yi-Hong Cai; Hsun Lee; Yu-Meng Ou; Chih-Hao Hsiao; Chien-Wei Tsao; Huan-Tsung Chang; Yi-Sheng Wang
Journal:  J Am Soc Mass Spectrom       Date:  2016-04-28       Impact factor: 3.109

2.  Matrix segregation as the major cause for sample inhomogeneity in MALDI dried droplet spots.

Authors:  Stefan J Gabriel; Clemens Schwarzinger; Bettina Schwarzinger; Ulrich Panne; Steffen M Weidner
Journal:  J Am Soc Mass Spectrom       Date:  2014-04-30       Impact factor: 3.109