Xiao-Zhu Lin1, Haruhiko Machida, Isao Tanaka, Rika Fukui, Eiko Ueno, Ke-Min Chen, Fu-Hua Yan.
Abstract
The purpose of this study is to compare CT images of the pancreas reconstructed with model-based iterative reconstruction (MBIR), adaptive statistical iterative reconstruction (ASiR), and filtered back projection (FBP) techniques for image quality and pancreatic duct (PD) depiction. Data from 40 patients with contrast-enhanced abdominal CT [CTDIvol: 10.3 ± 3.0 (mGy)] during the late arterial phase were reconstructed with FBP, 40% ASiR-FBP blending, and MBIR. Two radiologists assessed the depiction of the main PD, image noise, and overall image quality using 5-point scale independently. Objective CT value and noise were measured in the pancreatic parenchyma, and the contrast-to-noise ratio (CNR) of the PD was calculated. The Friedman test and post-hoc multiple comparisons with Bonferroni test following one-way ANOVA were used for qualitative and quantitative assessment, respectively. For the subjective assessment, scores for MBIR were significantly higher than those for FBP and 40% ASiR (all P < 0.001). No significant differences in CT values of the pancreatic parenchyma were noted among FBP, 40% ASiR, and MBIR images (P > 0.05). Objective image noise was significantly lower and CNR of the PD was higher with MBIR than with FBP and 40% ASiR (all P < 0.05). Our results suggest that pancreatic CT images reconstructed with MBIR have lower image noise, better image quality, and higher conspicuity and CNR of the PD compared with FBP and ASiR.Entities:
Mesh:
Substances:
Year: 2014 PMID: 24496703 DOI: 10.1007/s00261-014-0081-5