Begoña Molinero-Abad1, M Asunción Alonso-Lomillo2, Olga Domínguez-Renedo1, M Julia Arcos-Martínez1.
Abstract
Screen-printed carbon electrodes have been modified with tetrathiafulvalene and sulfite oxidase enzyme for the sensitive and selective detection of sulfite. Amperometric experimental conditions were optimized taking into account the importance of quantifying sulfite in wine samples and the inherent complexity of these samples, particularly red wine. The biosensor responds to sulfite giving a cathodic current (at +200 mV vs screen-printed Ag/AgCl electrode and pH 6) in a wide concentration range, with a capability of detection of 6 μM (α=β=0.05) at 60°C. The method has been applied to the determination of sulfite in white and red samples, with averages recoveries of 101.5% to 101.8%, respectively.Entities:
Keywords: Screen-printed carbon electrodes; Sulfite; Sulfite oxidase; Tetrathiafulvalene; Wine
Mesh:
Substances:
Year: 2014 PMID: 24491762 DOI: 10.1016/j.aca.2013.12.041
Source DB: PubMed Journal: Anal Chim Acta ISSN: 0003-2670 Impact factor: 6.558