Literature DB >> 24489018

Development of portable long counter with two different moderator materials.

Y Tanimura1, M Tsutsumi2, M Yoshizawa2.   

Abstract

A portable, light-weight long counter (LC) with small dimensions was developed. This LC consists of a (3)He thermal neutron counter, a cylindrical moderator and outer shields. It was designed to have an almost flat response in a neutron energy range of 0.4 eV to 5 MeV. The portable LC has a radius of 11 cm and a length of 39 cm. Its weight was successfully reduced to 15 kg. Polystyrene was employed instead of polyethylene for the front part of the moderator in order to increase the sensitivity to low-energy neutrons. The response function calculated using the MCNP code was consistent with the results of experiments using monoenergetic neutron calibration fields.
© The Author 2014. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24489018     DOI: 10.1093/rpd/nct345

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

1.  Characteristics of Protons Exiting from a Polyethylene Converter Irradiated by Neutrons with Energies between 1 keV and 10 MeV.

Authors:  D Nikezic; Mehrdad Shahmohammadi Beni; D Krstic; K N Yu
Journal:  PLoS One       Date:  2016-06-30       Impact factor: 3.240

  1 in total