J C Ramirez-San-Juan, R Ramos-Garcia, G Martinez-Niconoff, B Choi.
Abstract
One of the major constraints facing laser speckle imaging for blood-flow measurement is reliable measurement of the correlation time (τ(C)) of the back-scattered light and, hence, the blood's speed in blood vessels. In this Letter, we present a new model expression for integrated speckle contrast, which accounts not only for temporal integration but spatial integration, too, due to the finite size of the pixel of the CCD camera; as a result, we find that a correction factor should be introduced to the measured speckle contrast to properly determine τ(C); otherwise, the measured blood's speed is overestimated. Experimental results support our theoretical model.Entities: