Literature DB >> 24471147

Failure of a constrained acetabular liner without reinforcement ring disruption.

Jeffrey A Arthur, Derek F Amanatullah, Gannon D Kennedy, Paul E Di Cesare1.   

Abstract

Several risk factors for dislocation after total hip arthroplasty (THA) have been identified including operative-, patient-, and implant-related factors. The following case report describes the dislocation of a revision THA without disruption of the constrained liner or containment ring. The possible mechanisms leading to this type of failure include lever-out impingement and poor abductor function, or tension secondary to prior surgery. Dislocation without disruption of containment ring has not been described for the Pinnacle Acetabular Cup with the Enhanced Stability Constrained Liner (DePuy Orthopaedics, Warsaw, Indiana).

Entities:  

Mesh:

Year:  2013        PMID: 24471147

Source DB:  PubMed          Journal:  Am J Orthop (Belle Mead NJ)        ISSN: 1078-4519


  3 in total

1.  Letter to the Editor: Polyethylene Liner Dissociation is a Complication of the DePuy Pinnacle Cup: A Report of 23 Cases.

Authors:  Guillermo Bonilla; Maria Bautista
Journal:  Clin Orthop Relat Res       Date:  2016-03-14       Impact factor: 4.176

2.  Retentive Cup Arthroplasty in Selected Hip Fracture Patients-A Prospective Series With a Minimum 3-Year Follow-Up.

Authors:  Mustafa Yassin; Avraham Garti; Muhammad Khatib; Moshe Weisbrot; Dror Robinson
Journal:  Geriatr Orthop Surg Rehabil       Date:  2016-08-08

Review 3.  Causes of and treatment options for dislocation following total hip arthroplasty.

Authors:  Yian Lu; Haijun Xiao; Feng Xue
Journal:  Exp Ther Med       Date:  2019-07-03       Impact factor: 2.447

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.