Literature DB >> 24463104

Effect of different stapes prostheses on the passive vibration of the basilar membrane.

Monika Kwacz1, Piotr Marek2, Paweł Borkowski2, Wiktor Gambin3.   

Abstract

The effect of different stapes prostheses on the basilar membrane (BM) motion was determined. To that end, a three dimensional finite element (FE) model of the passive human cochlea was developed. Passive responses of the BM were found based on coupled fluid-structure interactions between the cochlear solid structures and the scala fluids. The passive BM vibrations in normal (healthy) cochlea were compared with vibrations in the cochlea in which a 0.4-mm piston or a proposed new type of prosthesis was implanted. The proposed chamber prosthesis was not experimentally implanted, but only numerically simulated. Design of the new chamber stapes prosthesis is presented for the first time in this paper. The simulation results showed 10-20 dB decrease in BM displacement amplitude in the case of the piston. In contrast, the BM responses in the cochlea with the new prosthesis are higher with respect to the healthy ear. The results obtained in this study are promising for further research to optimize the design of the new chamber stapes prosthesis.
Copyright © 2014 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Year:  2014        PMID: 24463104     DOI: 10.1016/j.heares.2014.01.004

Source DB:  PubMed          Journal:  Hear Res        ISSN: 0378-5955            Impact factor:   3.208


  4 in total

1.  Elastic Properties of the Annular Ligament of the Human Stapes--AFM Measurement.

Authors:  Monika Kwacz; Zygmunt Rymuza; Marcin Michałowski; Jarosław Wysocki
Journal:  J Assoc Res Otolaryngol       Date:  2015-06-04

Review 2.  Analytical and numerical modeling of the hearing system: Advances towards the assessment of hearing damage.

Authors:  Annalisa De Paolis; Marom Bikson; Jeremy T Nelson; J Alexander de Ru; Mark Packer; Luis Cardoso
Journal:  Hear Res       Date:  2017-02-02       Impact factor: 3.208

Review 3.  Design and optimization of auditory prostheses using the finite element method: a narrative review.

Authors:  Qianli Cheng; Han Yu; Junpei Liu; Qi Zheng; Yanru Bai; Guangjian Ni
Journal:  Ann Transl Med       Date:  2022-06

4.  New chamber stapes prosthesis - A preliminary assessment of the functioning of the prototype.

Authors:  Monika Kwacz; Magdalena Sołyga; Maciej Mrówka; Konrad Kamieniecki
Journal:  PLoS One       Date:  2017-05-23       Impact factor: 3.240

  4 in total