| Literature DB >> 24456988 |
Chang Ho1, Robert Oberle2, Isaac Wu2, Eugene Kim2.
Abstract
We compared image quality in 44 pediatric head computed tomographic exams reconstructed using four levels of blended iterative techniques (iDOSE 1-4) versus filtered back projection. Three radiologists reviewed the 220 series in a randomized, blinded manner. They scored each series based on gray-white matter differentiation, visibility of the pre-pontine cistern, caudate head, image noise, and image quality. Noise was measured in the vitreous and the thalamus. Composite scores were highest with iDOSE 4 (P<.0001). Interobserver agreement was fair to moderate. Spearman's rho of the vitreous (-0.62, P<.001) and thalamus (-0.58, P<.001) confirmed a significant inverse correlation between iDOSE level and noise.Entities:
Keywords: Dose reduction; Image quality; Iterative reconstruction; Pediatric head CT
Mesh:
Year: 2013 PMID: 24456988 DOI: 10.1016/j.clinimag.2013.12.009
Source DB: PubMed Journal: Clin Imaging ISSN: 0899-7071 Impact factor: 1.605