Automatic implantable cardioverter defibrillator/permanent pacemaker interaction: loss of pacemaker capture following AICD discharge.

M Slepian1, J H Levine, L Watkins, J Brinker, T Guarnieri.   

Abstract

A 78-year-old man treated with amiodarone for recurrent ventricular tachycardia, had sequential placement of a bipolar VVI pacemaker and an automatic implantable cardioverter defibrillator (AICD). During defibrillation threshold testing, there was failure to capture of the pacer in the post-shock period. The time of failure to capture appeared energy-related: the greater the energy delivered, the longer the failure to capture. Careful attention will be necessary in constructing combined AICD/pacemaker units.