Literature DB >> 24445092

Contrast-enhanced ultrasound using real-time contrast harmonic imaging in invasive breast cancer: comparison of enhancement dynamics with three different doses of contrast agent.

Ariel Saracco1, Botond K Szabó2, Peter Aspelin3, Karin Leifland4, Ervin Tánczos5, Brigitte Wilczek4, Rimma Axelsson3.   

Abstract

BACKGROUND: In the last few years new potential applications have been developed for contrast-enhanced ultrasound (CEUS) and the management of breast diseases, but there is still some debate concerning the optimal dose to evaluate breast lesions, especially as a diagnostic tool.
PURPOSE: To compare different CEUS doses of injected contrast agent in order to establish an optimal dose for the diagnosis of invasive breast cancer.
MATERIAL AND METHODS: In Group A we compared the bolus dose of 1.2 mL vs. 2.4 mL and in Group B we compared the bolus dose of 2.4 mL vs. 4.8 mL (26 and 25 invasive carcinomas, respectively). CEUS was performed in real-time contrast harmonic imaging (CHI) using a L9-3 MHz probe. All examinations were recorded in a contrast side/side imaging mode loop for 120 s. Wash-in and wash-out patterns of the contrast agent were analyzed with advanced US quantification software and kinetic curves were used for statistical analysis.
RESULTS: In Group B (2.4 mL vs. 4.8 mL), more and stronger correlation was found among kinetic parameters (area under the curve, P < 0.00001; lognormal model parameters, μ, P = 0.0007 and σ, P < 0.0001; mean transit time, P < 0.0001; model-based wash-out ratios, W21m, P = 0.0002; W50m, P = 0.0001; time-to-peak, P = 0.005) as compared to Group A (1.2 mL vs. 2.4 mL).
CONCLUSION: The optimal way to evaluate kinetic features of invasive breast tumors using real-time CEUS is with an injection of contrast agent of either 2.4 mL or 4.8 mL. © The Foundation Acta Radiologica 2014 Reprints and permissions: sagepub.co.uk/journalsPermissions.nav.

Entities:  

Keywords:  Contrast-enhanced ultrasound; doses; invasive breast cancer; kinetic analysis

Mesh:

Substances:

Year:  2014        PMID: 24445092     DOI: 10.1177/0284185114520860

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  2 in total

1.  Quantitative Multiparametric Breast Ultrasound: Application of Contrast-Enhanced Ultrasound and Elastography Leads to an Improved Differentiation of Benign and Malignant Lesions.

Authors:  Panagiotis Kapetas; Paola Clauser; Ramona Woitek; Georg J Wengert; Mathias Lazar; Katja Pinker; Thomas H Helbich; Pascal A T Baltzer
Journal:  Invest Radiol       Date:  2019-05       Impact factor: 6.016

2.  Comparison of transducers with different frequencies in breast contrast-enhanced ultrasound (CEUS) using SonoVue as contrast agent.

Authors:  Yong-Mei Wang; Wei Fan; Kai Zhang; Li Zhang; Zhen Tan; Rong Ma
Journal:  Br J Radiol       Date:  2016-05-04       Impact factor: 3.039

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.