| Literature DB >> 24431777 |
G Arun Kumar1, B Mahesh2, Dadu George2.
Abstract
The aim of the study was to compare the stress distribution around implant in different bone qualities of D1, D2, D3, and D4 with straight and angled abutments using three dimensional finite element analysis. A three dimensional finite element model of the premaxilla region, and two solid 4.3 × 10 mm implant, one with a straight abutment and the other with an angled abutment was done. Four distinctly different bone qualities of D1, D2, D3, and D4 were made. A static load of 178 N was applied at the centre of incisal edge along the long axis of each abutment. The maximum equivalent von Misses stress values around the implants were recorded. The distribution of stresses changed considerably with abutment angulation. As angulation increased from 0° to 15° the concentration of Von Misses stresses shifted to the cortical layer of bone on the facial side of the fixture. Although Von Misses stress increased in straight abutment as the bone quality changed from D1 to D4, it was more noticeable under the loading side of the angulated abutments. The high stresses induced through angled abutments at the cervical zone of the implant due to forces and moments could be a dominant factor that may aggravate the peri-implant bone loss or changes the existing peri-implantitis direction.Entities:
Keywords: Angled abutment; Bone quality; Implant
Year: 2013 PMID: 24431777 PMCID: PMC3792315 DOI: 10.1007/s13191-012-0242-6
Source DB: PubMed Journal: J Indian Prosthodont Soc ISSN: 0972-4052