Three-dimensional computed tomographic angiography of the liver for planning hepatic surgery: effect of low tube voltage and the iterative reconstruction algorithm on image quality.

Keiichi Honda1, Seitaro Oda, Kazuhiro Katahira, Hiroo Kajihara, Shoji Morishita, Maki Honbori, Kenji Kawata, Jiro Nasu, Daisuke Utsunomiya, Yoshinori Funama, Yasuyuki Yamashita.   

Abstract

OBJECTIVE: The aim of this study was to evaluate the effect of a low tube voltage technique and hybrid iterative reconstruction (HIR) on image quality at 3-dimensional computed tomographic angiography (3D-CTA) of the liver.
METHODS: Before hepatic surgery, we randomly assigned 60 patients (17 women, 43 men; mean ± SD age, 68.9 ± 10.1 years) who had undergone 3D-CTA to 1 of 2 protocols; 30 patients underwent scanning under the conventional 120-kilovolt (peak) protocol with filtered back projection (P1); and 30 patients, under an 80-kilovolt (peak) protocol with HIR (P2). The estimated effective radiation dose, computed tomographic attenuation, image noise, contrast-to-noise ratio, and figure of merit were calculated, and the visual image quality of 3D-CTA was scored on a 4-point scale.
RESULTS: The mean effective radiation dose was significantly lower under P2 than P1 (4.8 ± 1.2 vs 7.2 ± 1.5 mSv, P < 0.01). P1 and P2 did not significantly differ with respect to the image noise (10.5 ± 2.3 vs 9.9 ± 1.6 Hounsfield units; P = 0.46). Computed tomographic attenuation, contrast-to-noise ratio, figure of merit, and the visual scores for image quality were higher under P2 than P1 (P < 0.01).
CONCLUSIONS: The use of low tube voltage and HIR can yield significantly improved image quality at 3D-CTA of the liver.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24424561     DOI: 10.1097/RCT.0b013e3182a9a518

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  2 in total

1.  Investigation of the Utility and Safety of Dynamic Computed Tomography with Vasodilators.

Authors:  Jun Watanabe; Kanenori Endo; Yasutaka Tanaka; Keisuke Goto; Shoichi Urushibara; Tomohiro Osaki; Shigeru Tatebe; Seiichi Nakamura; Yasuaki Hirooka; Masahide Ikeguchi
Journal:  Yonago Acta Med       Date:  2020-01-24       Impact factor: 1.641

2.  CT evaluation of living liver donor: Can 100-kVp plus iterative reconstruction protocol provide accurate liver volume and vascular anatomy for liver transplantation with reduced radiation and contrast dose?

Authors:  Morikatsu Yoshida; Daisuke Utsunomiya; Masafumi Kidoh; Hideaki Yuki; Seitaro Oda; Shinya Shiraishi; Hidekazu Yamamoto; Yukihiro Inomata; Yasuyuki Yamashita
Journal:  Medicine (Baltimore)       Date:  2017-06       Impact factor: 1.889

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.