Biomechanical effects of bone-implant fitness and screw breakage on the stability and stress performance of the nonstemmed hip system.

Chun-Ming Chen1, Chih-Ting Cheng2, Cheng-Shou Lin3, Shang-Chih Lin4, Chien-Chung Chiang5, Chu-An Luo1, Ching-Shiow Tseng1.   

Abstract

BACKGROUND: Some nonstemmed hip systems have been developed to avoid stress shielding and aseptic loosening, which are major drawbacks of stemmed hip arthroplasty. Without the stem, the cup over the femoral head can be stabilized by anatomic fitness of the cup interior and mechanical fixation of the auxiliary screws.
METHODS: Using finite-element method, neck-shaped systems with two bone-cup fitness situations and four types of screw breakages are systematically investigated to evaluate their biomechanical effects on construct performances. The construct stresses and interfacial micromotion were chosen for comparison between two bone-cup fitness situations and four types of screw breakages.
FINDINGS: The screw breakage deteriorates the stresses of the mating screw and the neck cup and loosens the bone-cup interfaces. The breakages of central and locking screws decrease the bone stress by about 43.2% and 12.7%, respectively. This indicates that the central screw is a more effective load-bearer for the superimposed cup than the locking screw. As compared with the fitting cup, the stress of cup and the bone stresses of the unfitting cup obviously increase. This demonstrates that the load-transferring path at the cup bottom is important in directly relieving the prosthetic stresses.
INTERPRETATION: Any screw design inducing stress concentration should be validated to avoid screw breakage. Comparatively, surgical unfitness has a more significant effect on the construct performance than does the screw breakage. Even for custom-made cups, cautious preparation of the neck resection is still necessary to ensure intimate bone-cup contact.
Copyright © 2013. Published by Elsevier Ltd.

Entities:  

Keywords:  Finite-element; Hip nonstemmed; Neck cup; Screw breakage; Stress-shielding

Mesh:

Year:  2013        PMID: 24418222     DOI: 10.1016/j.clinbiomech.2013.11.017

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063