| Literature DB >> 24417433 |
J Burns1, S Hou, C B Riley, R A Shaw, N Jewett, J T McClure.
Abstract
BACKGROUND: Rapid, economical, and quantitative assays for measurement of camelid serum immunoglobulin G (IgG) are limited. In camelids, failure of transfer of maternal immunoglobulins has a reported prevalence of up to 20.5%. An accurate method for quantifying serum IgG concentrations is required.Entities:
Keywords: Camelid; Failure of passive transfer; Radial immunodiffusion
Mesh:
Substances:
Year: 2014 PMID: 24417433 PMCID: PMC4858009 DOI: 10.1111/jvim.12258
Source DB: PubMed Journal: J Vet Intern Med ISSN: 0891-6640 Impact factor: 3.333
Figure 1Representative infrared spectrum of alpaca serum. The strongest features arise from proteins with relatively less abundant serum components contributing relatively weak fingerprints. The absorption at 2,062 cm−1 originates with the KSCN − internal standard.
Figure 2A scatter plot comparing the immunoglobulin G (IgG) concentrations obtained from radial immunodiffusion (RID) and Fourier‐transform infrared spectroscopy (FTIR) methods. The asterisks denote samples used in building the calibration model and the circles indicate the samples in the test set. 15 partial least squares (PLS) factors were retained. If RID and FTIR give comparable results, the data points should distribute closely around the reference line of 45°.
Figure 3A Bland‐Altman plot of the differences in the immunoglobulin G (IgG) concentrations in the test set as obtained by radial immunodiffusion (RID) and Fourier‐transform infrared spectroscopy (FTIR) methods. The solid horizontal line represents the mean difference between RID and FTIR assays (−20.4 mg/dL) and the dashed lines represent the 95% confidence interval. If there is no systematic bias between RID and FTIR method, the mean value of the differences should be close to zero. If the model assumption is that the errors are independent and identically distributed and follow a normal distribution, the data points should distribute around the mean of the difference randomly. Narrow dispersion of the data points means low measurement uncertainties.
Figure 4Normal probability plot for the differences of the immunoglobulin G (IgG) concentrations in test set obtained from radial immunodiffusion (RID) and Fourier‐transform infrared spectroscopy (FTIR) methods. If the measurement errors follow a normal distribution, the data points should be largely located in the reference line.
Figure 5Coefficient of variance plots for the radial immunodiffusion (RID) and Fourier‐transform infrared spectroscopy (FTIR) methods. Lower coefficient of variance means high precision of the test methods; this comparison shows that the precision of the RID and FTIR methods are roughly comparable.
FTIR IgG false negative samples compared to IgG RID results identified within the entire data set (n = 175).
| Sample | Data Set | FTIR IgG Value (mg/dL) | RID IgG Value (mg/dL) |
|---|---|---|---|
| A | Calibration | 1,075 | 719 |
| B | Calibration | 1,192 | 773 |
| C | Test | 1,602 | 861 |
| D | Test | 1,307 | 822 |
FTIR, Fourier‐transform infrared spectroscopy; IgG, immunoglobulin G; RID, radial immunodiffusion.