| Literature DB >> 24417267 |
Md Azahar Ali1, S Srivastava, M K Pandey, Ved V Agrawal, R John, B D Malhotra.
Abstract
We propose a label-free biosensor platform to investigate the binding kinetics using antigen-antibody interaction via electrochemical and surface plasmon resonance (SPR) techniques. The L-cysteine in situ capped cadmium sulfide (CdS; size < 7 nm) quantum dots (QDs) self-assembled on gold (Au) coated glass electrode have been covalently functionalized with apolipoprotein B-100 antibodies (AAB). This protein conjugated QDs-based electrode (AAB/CysCdS/Au) has been used to detect lipid (low density lipoprotein, LDL) biomolecules. The electrochemical impedimetric response of the AAB/CysCdS/Au biosensor shows higher sensitivity (32.8 kΩ μM(-1)/cm(2)) in the detection range, 5-120 mg/dL. Besides this, efforts have been made to investigate the kinetics of antigen-antibody interactions at the CysCdS surface. The label-free SPR response of AAB/CysCdS/Au biosensor exhibits highly specific interaction to protein (LDL) with association constant of 33.4 kM(-1) s(-1) indicating higher affinity toward LDL biomolecules and a dissociation constant of 0.896 ms(-1). The results of these studies prove the efficacy of the CysCdS-Au platform as a high throughput compact biosensing device for investigating biomolecular interactions.Entities:
Mesh:
Substances:
Year: 2014 PMID: 24417267 DOI: 10.1021/ac403543g
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986