Literature DB >> 24416820

[Comparison of dynamic contour tonometry, Goldmann applanation Tonometry and tonopen for measuring intraocular pressure in normal tension glaucoma].

Shimon Kurtz1, Uri Soiberman1, Gabi Shemesh1.   

Abstract

BACKGROUND: Normal tension glaucoma (NTG) is a chronic progressive optic neuropathy, characterized by relatively normal intraocular pressure (IOP) measurements. This study utilized three modalities in the measurement of lOP: dynamic contour tonometry (Pascal-DCT), Goldmann applanation tonometry (GAT) and Tonopen XL.
OBJECTIVES: To compare the lOP measurements of the three devices in eyes with NTG.
METHODS: This retrospective study of thirty-three NTG patients (66 eyes) who underwent lOP measurements using the three devices was approved by a local ethics committee. The lOP measurements were stratified according to central corneal thickness (CCT) and corneal curvature (CC).
RESULTS: The mean DCT IOP was higher (16.22 +/- 3.04 mmHg) than the mean GAT IOP (13.14 +/- 2.83 mmHg, P < 0.0042). The mean Tonopen XL lOP was lower than the mean GAT IOP (12.76 +/- 3.11 mmHg, P < 0.001). GAT and DCT measurements were significantly infLuenced by corneal curvature (P = 0.004 and P < 0.0001, respectiveLy).
CONCLUSIONS: lOPs measured by GAT or Tonopen XL were consistently lower than by DCT. CC significantly influenced GAT and DCT results.

Entities:  

Mesh:

Year:  2013        PMID: 24416820

Source DB:  PubMed          Journal:  Harefuah        ISSN: 0017-7768


  1 in total

1.  Optimization of orbital retraction during endoscopic transorbital approach via quantitative measurement of the intraocular pressure - [SevEN 006].

Authors:  Woohyun Kim; Ju Hyung Moon; Eui Hyun Kim; Chang-Ki Hong; Jisang Han; Je Beom Hong
Journal:  BMC Ophthalmol       Date:  2021-02-08       Impact factor: 2.209