Literature DB >> 24412563

A gamma-gamma coincidence/anticoincidence spectrometer for low-level cosmogenic (22)Na/(7)Be activity ratio measurement.

Weihua Zhang1, Kurt Ungar2, Matthew Stukel2, Pawel Mekarski2.   

Abstract

In this study, a digital gamma-gamma coincidence/anticoincidence spectrometer was developed and examined for low-level cosmogenic (22)Na and (7)Be in air-filter sample monitoring. The spectrometer consists of two bismuth germanate scintillators (BGO) and an XIA LLC Digital Gamma Finder (DGF)/Pixie-4 software and card package. The spectrometer design allows a more selective measurement of (22)Na with a significant background reduction by gamma-gamma coincidence events processing. Hence, the system provides a more sensitive way to quantify trace amounts of (22)Na than normal high resolution gamma spectrometry providing a critical limit of 3 mBq within a 20 h count. The use of a list-mode data acquisition technique enabled simultaneous determination of (22)Na and (7)Be activity concentrations using a single measurement by coincidence and anticoincidence mode respectively.
Copyright © 2014 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  (22)Na; (7)Be; BGO detector; Gamma–Gamma coincidence/anticoincidence

Mesh:

Substances:

Year:  2014        PMID: 24412563     DOI: 10.1016/j.jenvrad.2013.12.018

Source DB:  PubMed          Journal:  J Environ Radioact        ISSN: 0265-931X            Impact factor:   2.674