| Literature DB >> 24388905 |
Yashan Wang1, Daohong Zhang1, Wei Liu1, Xiao Zhang1, Shaoxuan Yu1, Tao Liu1, Wentao Zhang1, Wenxin Zhu1, Jianlong Wang2.
Abstract
Existence of endotoxin in food and injection products indicates bacterial contaminations and therefore poses threat to human health. Herein, a simple and rapid colorimetric method for the effective detection of endotoxin in food and injections based on counterion-mediated gold nanorods aggregation is first proposed. By taking advantage of the color change of unmodified gold nanorods resulted from endotoxin mediated gold nanorods aggregation, endotoxin could be detected in the concentration range of 0.01-0.6 μM. Further, we studied the performance of gold nanorods with different aspect ratios (2.7 and 3.3) in determination of endotoxin and found that gold nanorods with higher aspect ratio (AR) showed superiority in the sensing sensitivity of endotoxin. A good specificity for endotoxin, a detection limit of 0.0084 μM and recoveries ranging from 84% to 109% in spiked food and injection samples are obtained with the colorimetric method. Results demonstrate that the present method provides a novel and effective approach for on-site screening of endotoxin in common products, which is beneficial for monitoring and reducing the risk of bacterial contaminations in food and injections production.Entities:
Keywords: Colorimetric detection; Endotoxin; Gold nanorods
Mesh:
Substances:
Year: 2013 PMID: 24388905 DOI: 10.1016/j.bios.2013.12.006
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618