CBCT with specification of imaging dose and CNR by anatomical volume of interest.

Del Leary1, James L Robar2.   

Abstract

PURPOSE: A novel method has been developed for volume of interest (VOI) cone-beam CT (CBCT) imaging using a 2.35 MV/Carbon target linac imaging beam line combined with dynamic multileaf collimator sequences.
METHODS: The authors demonstrate the concept of acquisition of multiple, separate imaging volumes, where volumes can be either completely separated or nested, and are associated with predetermined imaging dose and contrast-to-noise ratio (CNR) characteristics. Two individual MLC sequences were established in the planning system (Eclipse, Varian Medical) to collimate the beam according to a defined inner VOI (e.g., containing a target volume under image guidance) and an outer VOI (e.g., including surrounding landmarks or organs-at-risk). MLC sequences were interleaved as a function of gantry angle to produce a reconstructed CBCT image with nested VOIs. By controlling the ratio of inner-to-outer ratio of MLC segments (and thus Monitor Units) during acquisition, the relative dose and CNR in the two volumes can be controlled. Inner-to-outer ratios of 2:1 to 6:1 were examined.
RESULTS: The concept was explored using an anatomical head phantom to assess image quality. A geometric phantom was used to quantify absolute dose and CNR values for the various sequences. The authors found that the dose in the outer VOI decreased by a functional relationship dependent on the inner-to-outer sequence ratio, while the CNR varied by the square root of dose, as expected.
CONCLUSIONS: In this study the authors demonstrate flexibility in VOI CBCT by tailoring the imaging dose and CNR distribution in separate volumes within the patient anatomy. This would allow for high quality imaging of a target volume for alignment purposes, with simultaneous low dose imaging of the surrounding anatomy (e.g., for coregistration).

Entities:  

Mesh:

Year:  2014        PMID: 24387517     DOI: 10.1118/1.4855835

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Low Z target switching to increase tumor endothelial cell dose enhancement during gold nanoparticle-aided radiation therapy.

Authors:  Ross I Berbeco; Alexandre Detappe; Panogiotis Tsiamas; David Parsons; Mammo Yewondwossen; James Robar
Journal:  Med Phys       Date:  2016-01       Impact factor: 4.071

2.  Investigation of planar image quality for a novel 2.5 MV diamond target beam from a radiotherapy linear accelerator.

Authors:  Jennifer M Borsavage; Amanda Cherpak; James L Robar
Journal:  Phys Imaging Radiat Oncol       Date:  2020-11-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.