Qili Zhao1, Ming Wu2, Maosheng Cui3, Yanding Qin1, Jin Yu1, Mingzhu Sun1, Xin Zhao1, Xizeng Feng2.
Abstract
This paper presents a novel micropipette aspiration (MA) method based on a common pneumatic micro-injection system. This method is the first to quantify the influence of capillary effect on aspiration pressure using a balance pressure model, and in return, uses the capillary effect to quantify the aspiration pressure. Subsequently, the seal between the cell and the micropipette is detected to judge and exclude the ineffective MA attempts. The rationality of the balance pressure model is validated by the designed micropipette-filling experiments. Through applied to elasticity-determination of the cells with different sizes, the feasibility and versatility of this MA method are proved. With abilities to quantify aspiration pressures and detect the seam between the cell and the micropipette, our method is expected to advance the application of the commercial pneumatic injector in the MA of cells. Moreover, with the quantified volume of the liquid entering into the micropipette during MA process, our method also has a potential applicability to the study of the permeability of the cell membrane in the future.Mesh:
Year: 2013 PMID: 24387437 DOI: 10.1063/1.4832979