Literature DB >> 24376941

Basic tests in mammography as a tool in quality improvement.

Anna Kowalik1, Ewelina Konstanty2.   

Abstract

BACKGROUND: Mammography is a radiological diagnostic method which relies on an X-ray examination of breasts and is a process involving the use of low-dose amplitude-X-rays (usually around 0.7 mSv). Combining the use of small doses and high quality images requires extensive quality protocols, part of them being included in regulations adopted by the Minister of Health. AIM: The aim of this study was to check the usefulness and efficacy of selected quality tests associated with mammography. MATERIAL/
METHODS: The study was performed in the mammography service of the Greater Poland Cancer Centre in Poznan. Following equipment was used: densitometer, sensitometer, mammographic scales, electronic scales, thermometer, hygrometer, PMMA plates, Europhantom, screen film contact phantom, viewing boxes and magnifying glasses. The methods were based on basic mammography tests. Quality control in mammography demands: clean darkroom, marked and clean cassettes, clean viewing boxes with homogenous light.
RESULTS: The results of the "Development Process" test show that each sensitometer has to be used with an appropriate densitometer. Phantoms with abnormal structures cannot be used to "AEC System - Solidity exposure" test. "Compression - The force of compression" test may only be carried out with suitable scales and compressible material. Analysis of rejected films shows that the main reasons for rejection were wrong collimation and underexposure.
CONCLUSION: Every quality control in mammography provides essential information about the functioning of a laboratory. Apart from recommended standard sterility, it should be remembered that equipment should always be adjusted and repaired.

Entities:  

Keywords:  Basic tests; Densitometer; Mammography; Quality control; Sensitometer

Year:  2010        PMID: 24376941      PMCID: PMC3863229          DOI: 10.1016/j.rpor.2010.08.004

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  2 in total

1.  Mammography quality assurance from A to Z.

Authors:  D M Farria; L W Bassett; C Kimme-Smith; N DeBruhl
Journal:  Radiographics       Date:  1994-03       Impact factor: 5.333

2.  Breast cancer screening programmes in 22 countries: current policies, administration and guidelines. International Breast Cancer Screening Network (IBSN) and the European Network of Pilot Projects for Breast Cancer Screening.

Authors:  S Shapiro; E A Coleman; M Broeders; M Codd; H de Koning; J Fracheboud; S Moss; E Paci; S Stachenko; R Ballard-Barbash
Journal:  Int J Epidemiol       Date:  1998-10       Impact factor: 7.196

  2 in total
  6 in total

1.  Investment in radiotherapy infrastructure positively affected the economic status of an oncology hospital.

Authors:  Mirella Smigielska; Piotr Milecki
Journal:  Rep Pract Oncol Radiother       Date:  2012-05-18

Review 2.  Organizational, technical, physical and clinical quality standards for radiotherapy.

Authors:  Marta Bogusz-Czerniewicz; Daniel Kaźmierczak
Journal:  Rep Pract Oncol Radiother       Date:  2012-06-03

Review 3.  Patterns of care of radiotherapy in México.

Authors:  Adela Poitevin-Chacón; José Hinojosa-Gómez
Journal:  Rep Pract Oncol Radiother       Date:  2012-10-04

4.  Decentralisation of radiation therapy. Is it possible and beneficial to patients? Experience of the first 5 years of a satellite radiotherapy unit in the province of Tarragona, Spain.

Authors:  Meritxell Arenas; David Gomez; Sebastià Sabater; Angeles Rovirosa; Albert Biete; Jordi Colomer
Journal:  Rep Pract Oncol Radiother       Date:  2014-11-08

Review 5.  Identification of patient's requirements in quality management system in health care institutions.

Authors:  Daniel Kaźmierczak; Marta Bogusz-Czerniewicz
Journal:  Rep Pract Oncol Radiother       Date:  2011-12-16

6.  Clinical quality standards for radiotherapy.

Authors:  Marta Bogusz-Czerniewicz
Journal:  Contemp Oncol (Pozn)       Date:  2012-02-29
  6 in total