Proximity of pacemaker and implantable cardioverter-defibrillator leads to coronary arteries as assessed by cardiac computed tomography.

Benjamin J Pang1, Subodh B Joshi, Elaine H Lui, Mark A Tacey, Jeff Alison, Sujith K Seneviratne, James D Cameron, Harry G Mond.   

Abstract

INTRODUCTION: There have been rare case reports of damage to adjacent coronary arteries by screw-in pacemaker and implantable cardioverter-defibrillator (ICD) leads. Our aim was to assess the proximity of pacemaker and ICD leads to the major coronary anatomy using cardiac computed tomography (CT).
METHODS: Cardiac CT images were retrospectively analyzed to assess the spatial relationship of device lead tips to the major coronary anatomy.
RESULTS: Fifty-two right ventricular (RV) leads (17 apical, 35 nonapical) and 35 right atrial (RA) leads were assessed. Leads on the RV antero-septal junction (20 of 52) were close (median 4.7 mm) to, and orientated toward, the left anterior descending (LAD) coronary artery. RA leads in the anterior (26 of 35) and lateral (seven of 35) walls of the RA appendage were not close to (16.9 ± 7.7 mm and 18.9 ± 12.4 mm, respectively) and directed away from the right coronary artery. However, an RA lead adjacent to the superior border of the tricuspid valve was 4.3 mm from the right coronary artery and an RA lead on the medial wall of the RA appendage was 1.6 mm away from the aorta. An RV pacemaker lead in the lateral wall of the RV inlet was 3.4 mm from the right coronary artery.
CONCLUSIONS: In our cohort, a majority of RV leads were on the antero-septal junction and close to the overlying LAD coronary artery. RA leads adjacent to the tricuspid valve or on the medial RA appendage were in close proximity to the right coronary artery and aorta, respectively. ©2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  CT; coronary arteries; pacing lead implantation

Mesh:

Year:  2013        PMID: 24372320     DOI: 10.1111/pace.12330

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

Review 1.  Beware of the coronary arteries with implantable cardiac electronic devices.

Authors:  S Serge Barold; Benjamin J Pang; Harry G Mond
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2017-09

2.  Right coronary artery perforation by an active-fixation atrial pacing lead resulting in life-threatening tamponade.

Authors:  Eiichiro Nakagawa; Yukio Abe; Ryushi Komatsu; Takahiko Naruko; Akira Itoh
Journal:  J Arrhythm       Date:  2015-03-16

3.  A pacemaker lead indenting the wall of the left anterior descending artery.

Authors:  Łukasz M Gabryel; Klaudyna Bugla; Jacek Jafra; Jacek Gabryel
Journal:  Postepy Kardiol Interwencyjnej       Date:  2022-08-19       Impact factor: 1.065