[Study on accuracy of endoscopic polyp size measurement by disposable graduated biopsy forceps].

Ping Liu1, Xiu Zhang, Hui-ping Lin, Hei-jing Jin, Qiang Leng, Jin-hao Zhang, Yang Zhang, Hang Yao, Kun-lan Wu.   

Abstract

OBJECTIVE: To study the accuracy of endoscopic polyp size measurement by disposable graduated biopsy forceps (DGBF).
METHODS: Accurate gradation of 1 mm was made in the wire of disposable graduated biopsy forceps, which was used to measure the size of tumors under endoscopy. Fifty-eight polyps from 43 patients underwent endoscopy in our department from May to June 2013 were enrolled. Size of polyp was measured and compared among DGBF, routine estimation and direct measurement after resection. The accuracy of polyp size measurement was investigated by four colonoscopists who had finished at least 2000 procedures of colonoscopy.
RESULTS: The mean diameter of post-polypectomy measurement was (1.02±0.84) cm. Diameter was less than 1 cm in 36 polyps, 1 to 2 cm in 15, and over 2 cm in 7. The mean diameter of visual estimation was (1.29±1.07) cm, and the difference was significant as compared with actual size (P=0.000). The mean diameter measured by DGBF was (1.02±0.82) cm, and the difference was not significant as compared with actual size (P=0.775). The ratio of visual estimation to actual size was 1.29±0.31, and DGBF estimation to actual size was 1.02±0.11 with significant difference (P=0.000). The accurate rate of DGBF in estimating polyp size was 77.6% (45/58), which was obviously higher as compared to visual estimation [19.0% (11/58), P=0.000].
CONCLUSION: The accuracy of DGBF as a scale in the estimation of poly size increases as compared to visual estimation.

Entities:  

Mesh:

Year:  2013        PMID: 24369393

Source DB:  PubMed          Journal:  Zhonghua Wei Chang Wai Ke Za Zhi        ISSN: 1671-0274


  1 in total

1.  Measurement system that improves the accuracy of polyp size determined at colonoscopy.

Authors:  Qiang Leng; Hei-Ying Jin
Journal:  World J Gastroenterol       Date:  2015-02-21       Impact factor: 5.742