| Literature DB >> 24364899 |
Stanley Tay1, Brian Spain, Kirstie Morandell, Jesse Gilson, Laurence Weinberg, David Story.
Abstract
BACKGROUND: There are wide variations in the physical designs and attributes between different brands of intravenous cannulae that makes product selection and purchasing difficult. In a systematic assessment to guide purchasing, we assessed two cannulae - Cannula P and I. We proposed that the results of in-vitro performance testing of the cannulae would be associated with preference after clinical comparison.Entities:
Year: 2013 PMID: 24364899 PMCID: PMC3882495 DOI: 10.1186/1471-2253-13-49
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.217
Figure 1Setup up to indirectly measure resistance by recording the pressure generated through an intravenous cannula’s needle at various flow rates. Consisting of A: intravenous cannula; B: 3-way tap; C: 50 ml syringe; D: Pressure transducer; E: Syringe driver.
Figure 2Setup up to indirectly measure static friction by recording the force required to displace the catheter away from the needle. Consisting of A: intravenous cannula; B: electronic spring scale.
Figure 3Pressure (mmHg) measured for a given flow rate (L.hr) as an indirect measure of resistance in Cannula (•) P and () I: 22G (A), 20G (B) and 18G (C). Only the mean pressure from the 10 measurements at each designated flow rates are plotted. Standard deviations are represented by error bars. Trends lines are also drawn for Cannula () P and () I.
Measurements of resistance through the needle in Cannula P and I
| 22G resistance; mmHg.L-1.hr-1 | 728 (3.7) | 421 (3.6) | 307 [289 to 325] | 2 × 10-14 |
| 20G resistance; mmHg.L-1.hr-1 | 287 (6.3) | 152 (3.1) | 135 [125 to 144] | 3 × 10-10 |
| 18G resistance; mmHg.L-1.hr-1 | 71 (2.4) | 44 (4.7) | 27 [26 to 28] | 1 × 10-16 |
Values are mean with (relative standard deviation %) or [95% confidence interval].
Measurements of force required to displace the catheter from the needle in Cannula P and I as an indirect measure of static resistance
| 22G static resistance; N | 1.80 (25.2) | 0.39 (28.7) | 1.41 [1.09 to 1.73] | 2 × 10-6 |
| 20G static resistance; N | 0.73 (40.0) | 0.55 (37.3) | 0.19 [-0.04 to 0.41] | 0.12 |
| 18G static resistance; N | 2.40 (37.4) | 0.44 (13.3) | 1.96 [1.40 to 2.52] | 7 × 10-5 |
Values are mean with (relative standard deviation %) or [95% confidence interval].
Survey responses from consultant anaesthetists regarding the use of Cannula P and I
| Have you used both Cannula P and I? (n = 17) | |
| Yes | 16 (94%) |
| No | 1 (6%) |
| How does Cannula P compare to Cannula I? (n = 16) | |
| Significantly better | 0 (0%) |
| Better | 0 (0%) |
| No difference | 0 (0%) |
| Worse | 7 (44%) |
| Significantly worse | 9 (56%) |
| What should we do? (n = 17) | |
| Use Cannula P over I | 0 (0%) |
| Use either Cannula P or I | 1 (6%) |
| Use Cannula I over P | 13 (76%) |
| Stop using both Cannula P and I | 1 (6%) |
| Stop using all safety cannulae | 2 (12%) |
Manufacturer specified physical characteristics of catheters
| 18G (1.3 mm) | Length (mm) | 32 | 32 |
| Flow rate (ml.min-1) | 95 | 105 | |
| 20G (1.1 mm) | Length (mm) | 32 | 32 |
| Flow rate (ml.min-1) | 65 | 60 | |
| 22G (0.9 mm) | Length (mm) | 25 | 25 |
| Flow rate (ml.min-1) | 36 | 35 |