Wenjing Wei1, Yilin Song2, Wentao Shi2, Nansen Lin2, Tingjun Jiang1, Xinxia Cai3.
Abstract
The mammalian central nervous system (CNS) relies on a constant supply of external glucose for its undisturbed operation. This article presents an implantable Multi-Electrode Array (MEA) probe for brain glucose measurement. The MEA was implemented on Silicon-On-Insulator (SOI) wafer using Micro-Electro-Mechanical-Systems (MEMS) methods. There were 16 platinum recording sites on the probe and enzyme glucose oxidase (GOx) was immobilized on them. The glucose sensitivity of the MEA probe was as high as 489 µA mM(-1) cm(-2). 1,3-Phenylenediamine (mPD) was electropolymerized onto the Pt recording surfaces to prevent larger molecules such as ascorbic acid (AA), 3,4-dihydroxyphenylacetic acid (DOPAC), serotonin (5-HT), and dopamine (DA) from reaching the recording sites surface. The MEA probe was implanted in the anesthetized rat striatum and responded to glucose levels which were altered by intraperitoneal injection of glucose and insulin. After the in vivo experiment, the MEA probe still kept sensitivity to glucose, these suggested that the MEA probe was reliable for glucose monitoring in brain extracellular fluid (ECF).Entities:
Keywords: Glucose biosensor; MEMS; Microelectrode; Rat brain; Real-time monitoring
Mesh:
Substances:
Year: 2013 PMID: 24362080 DOI: 10.1016/j.bios.2013.11.048
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618