Literature DB >> 24361117

Performance of two strategies for urgent ANCA and anti-GBM analysis in vasculitis.

Anoek A E de Joode1, Caroline Roozendaal2, Marcel J van der Leij2, Laura B Bungener2, Jan Stephan F Sanders3, Coen A Stegeman3.   

Abstract

BACKGROUND: In anti-neutrophil cytoplasmic antibodies (ANCA) associated small vessel vasculitis (AAV), rapid testing for ANCA and anti-glomerular basement membrane (GBM) antibodies may be beneficial for therapeutic purpose.
OBJECTIVE: We analysed the diagnostic performance of two rapid ANCA and anti-GBM test methods in 260 patients with suspected AAV.
METHODS: Between January 2004 and November 2010, we analysed 260 samples by qualitative Dotblot (Biomedical Diagnostics); retrospective analysis followed with directly coated highly sensitive automated Phadia ELiA and ELiA anti-GBM. Results were related to the final clinical diagnosis and compared with routine capture ELISA.
RESULTS: Seventy-four patients had a final diagnosis of AAV (n=62) or anti-GBM disease (n=12). Both Dotblot and ELiA detected all 12 cases of anti-GBM disease; 2 false positive results were found. Dotblot detected ANCA in 56 of 62 AAV patients (sensitivity 90%, NPV 97%), and showed 5 false positives (specificity 97%, PPV 90%). The Phadia ELiA anti-PR3(s) or anti-MPO(s) was positive in 57 of 62 AAV patients (sensitivity 92%, NPV 97%), and had 5 false positives (specificity 97%, PPV 88%). Routine capture ELISA was equally accurate (sensitivity 94%, specificity 97%, PPV 88%, NPV 98%).
CONCLUSION: The Dotblot and Phadia ELiA on anti-GBM, anti-PR3(s) and anti-MPO(s) performed excellently; results were almost identical to routine ELISA. When suspicion of AAV or anti-GBM disease is high and diagnosis is urgently needed, both tests are very powerful for rapid serological diagnosis. Further studies have to confirm the test performances in samples routinely presented for ANCA testing and in follow-up of positive patients.
Copyright © 2013 European Federation of Internal Medicine. Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Dotblot; ELISA; ELiA; Indirect immunofluorescence; Serology; Systemic vasculitis

Mesh:

Substances:

Year:  2013        PMID: 24361117     DOI: 10.1016/j.ejim.2013.11.011

Source DB:  PubMed          Journal:  Eur J Intern Med        ISSN: 0953-6205            Impact factor:   4.487


  6 in total

Review 1.  Anti-neutrophil cytoplasmic antibodies and their clinical significance.

Authors:  Supaporn Suwanchote; Muanpetch Rachayon; Pongsawat Rodsaward; Jongkonnee Wongpiyabovorn; Tawatchai Deekajorndech; Helen L Wright; Steven W Edwards; Michael W Beresford; Pawinee Rerknimitr; Direkrit Chiewchengchol
Journal:  Clin Rheumatol       Date:  2018-03-10       Impact factor: 2.980

2.  Automation, consolidation, and integration in autoimmune diagnostics.

Authors:  Renato Tozzoli; Federica D'Aurizio; Danilo Villalta; Nicola Bizzaro
Journal:  Auto Immun Highlights       Date:  2015-07-03

3.  The kidneys and ANCA-associated vasculitis: from pathogenesis to diagnosis.

Authors:  Olumide Olatubosun Rowaiye; Mariusz Kusztal; Marian Klinger
Journal:  Clin Kidney J       Date:  2015-04-05

4.  M2 macrophage is the predominant phenotype in airways inflammatory lesions in patients with granulomatosis with polyangiitis.

Authors:  Alexandre Wagner Silva de Souza; Mirjan van Timmeren; Jan-Stephan Sanders; Coen Stegeman; Peter Heeringa; Cees G M Kallenberg; Johanna Westra
Journal:  Arthritis Res Ther       Date:  2017-05-18       Impact factor: 5.156

5.  Chemiluminescent immunoassay technology: what does it change in autoantibody detection?

Authors:  Luigi Cinquanta; Desré Ethel Fontana; Nicola Bizzaro
Journal:  Auto Immun Highlights       Date:  2017-06-24

6.  Falsely positive anti-glomerular basement membrane antibodies in a patient with hantavirus induced acute kidney injury - a case report.

Authors:  H W Zijlstra; A H L Mulder; F Geeraedts; F Visser
Journal:  BMC Nephrol       Date:  2018-10-22       Impact factor: 2.388

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.