Literature DB >> 24360820

Comparison of circuit patency and exchange rates between 2 different continuous renal replacement therapy machines.

Seyed Amirhossein Razavi1, Mary D Still2, Sharon J White3, Timothy G Buchman4, Michael J Connor5.   

Abstract

BACKGROUND: Continuous renal replacement therapy (CRRT) is an important tool in the care of critically ill patients. However, the impact of a specific CRRT machine type on the successful delivery of CRRT is unclear. The purpose of this study was to evaluate the effectiveness of CRRT delivery with an intensive care unit (ICU) bedside nurse delivery model for CRRT while comparing circuit patency and circuit exchange rates in 2 Food and Drug Administration-approved CRRT devices. This article presents the data comparing circuit exchange rates for 2 different CRRT machines.
MATERIALS AND METHODS: A group of ICU nurses were selected to undergo expanded training in CRRT operation and empowered to deliver all aspects of CRRT. The ICU nurses then provided all aspects of CRRT on 2 Food and Drug Administration-approved CRRT devices for 6 months. Each device was used exclusively in the designated ICU for a 2-week run-in period followed by 3-month data collection period. The primary end point for the study was the differences in average number of filter exchanges per day during each CRRT event.
RESULTS: A total of 45 unique patients who underwent 64 separate CRRT treatment periods were included. Four CRRT events were excluded (see text for details). Twenty-eight CRRT events occurred in the NxStage System One arm (NxStage Medical, Lawrence, Mass) and 32 events in the Gambro Prismaflex arm (Gambro Renal Products, Boulder, Colo). Average (SD) filter exchanges per day was 0.443 (0.60) for the NxStage System One machine and 0.553 (0.65) for Gambro Prismaflex machine (P = .09).
CONCLUSIONS: There was no demonstrable difference in circuit patency as defined by the rate of filter exchanges per day of CRRT therapy.
Copyright © 2014 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Circuit survival; Continuous renal replacement therapy; Filter exchange; Gambro Prismaflex machine; NxStage System One machine; Quality improvement

Mesh:

Substances:

Year:  2013        PMID: 24360820     DOI: 10.1016/j.jcrc.2013.11.012

Source DB:  PubMed          Journal:  J Crit Care        ISSN: 0883-9441            Impact factor:   3.425


  1 in total

1.  Anticoagulation Strategies and Filter Life in COVID-19 Patients Receiving Continuous Renal Replacement Therapy: A Single-Center Experience.

Authors:  Divya Shankaranarayanan; Thangamani Muthukumar; Tarek Barbar; Aarti Bhasin; Supriya Gerardine; Perola Lamba; Lorenz Leuprecht; Sanjay P Neupane; Thalia Salinas; Daniil Shimonov; Elly Varma; Frank Liu
Journal:  Clin J Am Soc Nephrol       Date:  2020-09-17       Impact factor: 8.237

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.