Scapular notching in reverse shoulder arthroplasty: validation of a computer impingement model.

Christopher P Roche, Yann Marczuk, Thomas W Wright, Pierre-Henri Flurin, Sean G Grey, Richard B Jones, Howard D Routman, Gregory J Gilot, Joseph D Zuckerman.   

Abstract

PURPOSE: The purpose of this study is to validate a reverse shoulder computer impingement model and quantify the impact of implant position on scapular impingement by comparing it to that of a radiographic analysis of 256 patients who received the same prosthesis and were followed postoperatively for an average of 22.2 months.
METHODS: A geometric computer analysis quantified anterior and posterior scapular impingement as the humerus was internally and externally rotated at varying levels of abduction and adduction relative to a fixed scapula at defined glenoid implant positions. These impingement results were compared to radiographic study of 256 patients who were analyzed for notching, glenoid baseplate position, and glenosphere overhang.
RESULTS: The computer model predicted no impingement at 0° humeral abduction in the scapular plane for the 38 mm, 42 mm, and 46 mm devices when the glenoid baseplate cage peg is positioned 18.6 mm, 20.4 mm, and 22.7 mm from the inferior glenoid rim (of the reamed glenoid) or when glenosphere overhang of 4.6 mm, 4.7 mm, and 4.5 mm was obtained with each size glenosphere, respectively. When compared to the radiographic analysis, the computer model correctly predicted impingement based upon glenoid base- plate position in 18 of 26 patients with scapular notching and based upon glenosphere overhang in 15 of 26 patients with scapular notching.
CONCLUSIONS: Reverse shoulder implant positioning plays an important role in scapular notching. The results of this study demonstrate that the computer impingement model can effectively predict impingement based upon implant positioning in a majority of patients who developed scapular notching clinically. This computer analysis provides guidance to surgeons on implant positions that reduce scapular notching, a well-documented complication of reverse shoulder arthroplasty.

Entities:  

Mesh:

Year:  2013        PMID: 24344620

Source DB:  PubMed          Journal:  Bull Hosp Jt Dis (2013)        ISSN: 2328-4633


  4 in total

1.  Clinical implications of scapular notching at 2 and 5-year follow-up after reverse total shoulder arthroplasty.

Authors:  Robert J Shelley; Mikalyn T DeFoor; Stephen A Parada; Lynn A Crosby
Journal:  J Orthop       Date:  2020-08-19

Review 2.  Reverse Shoulder Arthroplasty Biomechanics.

Authors:  Christopher P Roche
Journal:  J Funct Morphol Kinesiol       Date:  2022-01-19

3.  The influence of glenoid component position on scapular notching: a detailed radiographic analysis at midterm follow-up.

Authors:  Nicholas C Duethman; William R Aibinder; Ngoc Tram V Nguyen; Joaquin Sanchez-Sotelo
Journal:  JSES Int       Date:  2020-01-14

4.  Stable glenoid component of reverse total shoulder arthroplasty at 2 years as measured with model-based radiostereometric analysis (RSA).

Authors:  Alexander Nilsskog Fraser; Berte Bøe; Tore Fjalestad; Jan Erik Madsen; Stephan M Röhrl
Journal:  Acta Orthop       Date:  2021-07-01       Impact factor: 3.717

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.