Prevalence of detectable venous pressure drops expected with venous needle dislodgement.

Werner Ribitsch1, Gernot Schilcher, Hildegard Hafner-Giessauf, Peter Krisper, Jörg H Horina, Alexander R Rosenkranz, Daniel Schneditz.   

Abstract

Venous needle dislodgement (VND) is a potentially fatal complication during hemodialysis (HD) treatment and the venous pressure monitor is the most widely used device for its detection. VND can only be detected by the venous sensor if the resulting pressure drop exceeds the difference between the actual venous pressure and the lower alarm limit. In clinical practice, the lower alarm limit is usually set 30-40 mmHg below the actual venous pressure to avoid a disproportionate high number of nuisance alarms. The aim of this study was to quantify the number of fistulas and grafts in a group of HD patients where venous pressure monitoring can be expected to detect VND. We determined intra-access pressures in 99 chronic HD patients. Sixty-five (65.7%) had a fistula and 34 (34.3%) had a prosthetic graft as a vascular access. Mean intra-access pressure (Pa ) in fistulas was 32.6 ± 23.5 mmHg, whereas in grafts mean Pa was 60.9 ± 19.5 mmHg. Nineteen (29.2%) of the fistulas and 32 (94.1%) of the grafts exhibited an intra-access pressure above 40 mmHg. Therefore, in our study nearly all grafts but only 29% of fistulas would fulfill the requirement for venous pressure monitoring to detect VND.
© 2013 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2013        PMID: 24341865     DOI: 10.1111/sdi.12169

Source DB:  PubMed          Journal:  Semin Dial        ISSN: 0894-0959            Impact factor:   3.455


  4 in total

1.  Renal Association Clinical Practice Guideline on Haemodialysis.

Authors:  Damien Ashby; Natalie Borman; James Burton; Richard Corbett; Andrew Davenport; Ken Farrington; Katey Flowers; James Fotheringham; R N Andrea Fox; Gail Franklin; Claire Gardiner; R N Martin Gerrish; Sharlene Greenwood; Daljit Hothi; Abdul Khares; Pelagia Koufaki; Jeremy Levy; Elizabeth Lindley; Jamie Macdonald; Bruno Mafrici; Andrew Mooney; James Tattersall; Kay Tyerman; Enric Villar; Martin Wilkie
Journal:  BMC Nephrol       Date:  2019-10-17       Impact factor: 2.388

Review 2.  Diagnosis, Treatment, and Prevention of Hemodialysis Emergencies.

Authors:  Manish Saha; Michael Allon
Journal:  Clin J Am Soc Nephrol       Date:  2016-11-09       Impact factor: 8.237

3.  Novel Wearable Device for Blood Leakage Detection during Hemodialysis Using an Array Sensing Patch.

Authors:  Yi-Chun Du; Bee-Yen Lim; Wei-Siang Ciou; Ming-Jui Wu
Journal:  Sensors (Basel)       Date:  2016-06-09       Impact factor: 3.576

4.  A device for surveillance of vascular access sites for bleeding: results from a clinical evaluation trial.

Authors:  Chang Yin Chionh; Desilyn Yuqing Soh; Chee How Tan; Jien-Yi Khaw; Ying Ching Wong; Shaohui Foong
Journal:  Sci Rep       Date:  2020-10-23       Impact factor: 4.379

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.