| Literature DB >> 24341860 |
Yuntong Zhang, Xue Zhao, Yang Tang, Chuncai Zhang, Shuogui Xu1, Yang Xie.
Abstract
BACKGROUND: Our preliminary retrospective study assessed outcomes after the use of Ni-Ti arched shape-memory connector (ASC) combined with partially threaded cancellous screws (PTCS) to repair coronal plane supracondylar-condylar femoral fractures.Entities:
Mesh:
Substances:
Year: 2013 PMID: 24341860 PMCID: PMC3878495 DOI: 10.1186/1471-2474-14-355
Source DB: PubMed Journal: BMC Musculoskelet Disord ISSN: 1471-2474 Impact factor: 2.362
Summary of the patient profiles and subsequent management with outcomes
| 1 | 31/F | fall | / | ASC×3&PTCS×2(6.5 mm) | 0-100 | 83/good |
| 2 | 38/M | Traffic accident | Tibial plateau fracture/medial meniscus tears | ASC×3&PTCS×1(6.5 mm) | 5-100 | 79/good |
| 3 | 39/M | Traffic accident | / | ASC×2&PTCS×2(4.5 mm) | 0-120 | 88/excelent |
| 4 | 32/M | fall | Calcaneus and patellar fracture / MCL injury | ASC×3&PTCS×2(6.5 mm) | 0-105 | 90/excelent |
| 5 | 34/M | fall | / | ASC×2&PTCS×1(4.5 mm) | 5-95 | 84/good |
| 6 | 35/M | fall | / | ASC×2&PTCS×1(6.5 mm) | 0-105 | 82/good |
| 7 | 44/M | Traffic accident | Tibial plateau fracture/MCL injury/ medial meniscus tears | ASC×3&PTCS×2(4.5 mm) | 0-115 | 92/excelent |
| 8 | 28/F | Traffic accident | Patellar fracture | ASC×2&PTCS×2(6.5 mm) | 0-95 | 88/excelent |
| 9 | 22/M | Traffic accident | MCL injury | ASC×2&PTCS×1(6.5 mm) | 0-110 | 93/excelent |
| 10 | 33/M | Traffic accident | medial meniscus tears/MCL injury | ASC×3&PTCS×2(4.5 mm) | 0-90 | 65/fair |
| 11 | 39/M | fall | / | ASC×2&PTCS×2(4.5 mm) | 0-95 | 90/excelent |
| 12 | 32/F | Traffic accident | Patellar fracture | ASC×3&PTCS×2(6.5 mm) | 0-90 | 80/good |
| 13 | 30/M | Traffic accident | / | ASC×2&PTCS×1(6.5 mm) | 0-85 | 68/fair |
| 14 | 43/F | Traffic accident | / | ASC×3&PTCS×2(4.5 mm) | 5-115 | 90/excelent |
| 15 | 29/M | Traffic accident | / | ASC×2&PTCS×1(6.5 mm) | 0-105 | 84/good |
| 16 | 30/M | Traffic accident | MCL injury | ASC×3&PTCS×2(4.5 mm) | 5-95 | 82/good |
| 17 | 38/M | Traffic accident | Patellar fracture | ASC×2&PTCS×1(6.5 mm) | 5-110 | 82/good |
| 18 | 34/M | fall | / | ASC×3&PTCS×2(6.5 mm) | 5-100 | 78/good |
| 19 | 38/F | fall | Patellar fracture | ASC×3&PTCS×2(6.5 mm) | 0-100 | 84/good |
| 20 | 34/M | Traffic accident | / | ASC×3&PTCS×2(6.5 mm) | 5-115 | 92/excelent |
| 21 | 33/M | Traffic accident | / | ASC×3&PTCS×2(4.5 mm) | 5-110 | 82/good |
ROM: range of motion, ASC: Ni-Ti arched shape-memory connector, PTCS: partially threaded cancellous screws.
Figure 1Ni-Ti arched shape-memory connectors.
Figure 2The device consists of compression arms that are connected to the waist.
Figure 3Intraoperative picture of ASC for fixation of comminuted supercondylar fractures.
Figure 4Representative case. (a,b) a 32-year-old male patient was presented with comminuted supracondylar-condylar fracture of the right distal femur, associated with patellar fracture and medial collateral ligament injury. (c) The CT shows coronal plane fracture of supracondylar-condylar femur. (d,e) Post-operation radiographs with 30 months follow-up indicate the fracture has healed well. No evidence of osteonecrosis or development of traumatic arthritis can be observed.