Spatiotemporal accuracy of real-time 3D echocardiography in the neonatal and pediatric setting--validation studies using small dynamic test objects.

U Herberg1, C Klebach1, J Faller1, H G Trier2, J Breuer1.   

Abstract

UNLABELLED: The precision of real-time 3D-echocardiography (RT3DE) is not sufficiently validated for small, fast-moving structures such as the neonatal and pediatric heart.
PURPOSE: To assess the spatiotemporal accuracy of RT3DE in small, moving test objects.
MATERIALS AND METHODS: Small, calibrated test objects in the size of neonatal and pediatric heart chambers were made from polyurethane foam or metal wire mesh and moved in a water bath through a calibrated dynamic test system. Using matrix transducers (X7-2, ie33 and X4-1, Sonos 7500, Philips, Andover, USA), 2 D and live 3 D datasets under variation of the motion speed (0.033 - 0.133 m/s corresponding to 50 - 200 heart cycles/minute), the volume rate and transducer position were recorded and analyzed (QLab 7.0, Philips).
RESULTS: 3 D datasets of the moving test objects showed relevant spatial distortion, which was obviously related to the sequential scanning technology of the matrix transducer. Different segments of a test object were not recorded simultaneously, but rather row-by-row, so that there was a time delay between the first and the last-recorded voxel of a single 3 D volume (mean±SD: 28.9 ± 7.82 m/s or 80 ± 7 % of the time duration of a 3 D volume). With increasing motion speed of the test object and reduced 3 D volume rate, the distortion artifacts increased significantly.
CONCLUSION: 3 D acquisitions using matrix technology demonstrate relevant spatiotemporal inaccuracies. This may lead to misinterpretations during the evaluation of the synchronicity of valvular or ventricular motion and incorrect definition of volume estimations. In particular, at higher heart rates and higher rates of movement, these limitations have to be taken into account in clinical practice. © Georg Thieme Verlag KG Stuttgart · New York.

Entities:  

Mesh:

Year:  2013        PMID: 24338695     DOI: 10.1055/s-0033-1355764

Source DB:  PubMed          Journal:  Ultraschall Med        ISSN: 0172-4614            Impact factor:   6.548


  2 in total

1.  Ventricular pressure-volume loops obtained by 3D real-time echocardiography and mini pressure wire-a feasibility study.

Authors:  Ulrike Herberg; Eva Gatzweiler; Thomas Breuer; Johannes Breuer
Journal:  Clin Res Cardiol       Date:  2013-02-09       Impact factor: 5.460

Review 2.  4D imaging of fetal right ventricle-feasibility study and a review of the literature.

Authors:  M Pasieczna; J Duliban; A Grzyb; J Szymkiewicz-Dangel
Journal:  Int J Cardiovasc Imaging       Date:  2021-09-20       Impact factor: 2.357

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.